Emerson 5081FG instruction manual Handling the Analyzer System Considerations

Page 27

Model 5081FG

Instruction Manual

IB-106-5081, Rev. 1.0

May 2005

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HART

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODEL 275/375

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HAND HELD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INTERFACE

 

TWO-WIRE IN SITU

O2

mV

 

4-20 mA OUTPUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SIGNAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(TWISTED PAIR)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OXYGEN ANALYZER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODEL 5081

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TRANSMITTER

 

 

 

 

 

 

 

+

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CALIBRATION CHECK

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GAS LINE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

24 VDC

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REFERENCE

TEMPERATURE

 

 

 

 

 

 

 

POWER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AIR LINE

TERMINATION IN

SUPPLY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

mV SIGNAL

 

 

 

INTRINSIC

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONTROL ROOM

 

 

 

ASSET MANAGEMENT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAFETY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BARRIER

 

 

 

SOLUTIONS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(OPTIONAL)

37240002

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 1-2. Two-Wire In Situ Oxygen Analyzer HART Connections and AMS Application

 

e. Handling the Analyzer

 

 

 

 

 

 

 

 

 

f.

System Considerations

The probe was specially packaged to pre- vent breakage due to handling. Do not re- move the padding material from the probe until immediately before installation.

It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to pre- vent possible equipment damage.

The oxygen probe is designed for in- dustrial applications. Treat with care to avoid physical damage. The probe contains components made from ce- ramic, which are susceptible to shock when mishandled. THE WARRANTY DOES NOT COVER DAMAGE FROM MISHANDLING.

Prior to installing your Rosemount Analytical Two-Wire In Situ Oxygen Analyzer, make sure you have all the components neces- sary to make the system installation. Ensure all the components are properly integrated to make the system functional.

After verifying that you have all the compo- nents, select mounting locations and deter- mine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considera- tions, convenience, and serviceability. Fig- ure 1-2 shows a typical system wiring. A typical system installation is illustrated in Figure 1-3.

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications 1-5

Image 27
Contents Model 5081FG Essential Instructions Highlights of Changes Effective May, 2005 RevSummary Page Model 5081FG Table of ContentsList of Illustrations List of Tables Model 5081FG Definitions PrefaceModel 5081FG Belangrijk Vigtigt Model 5081FG Tärkeää Model 5081FG Wichtig Importante Viktig Model 5081FG Model 5081FG Viktigt Model 5081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Description System ConfigurationSystem Overview ScopeSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOperating Load Region Lift OFF Maximum Power Supply VoltageOhms VDCInspect Section InstallationPRE-INSTALLATION Mechanical InstallationSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 5081FG Installing Model 5081 Transmitter Mounting Holes Cover LockCircuit Terminal END Block TBElectrical Installation Display Positioning AssemblyModel 5081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Calibration Check Gas Pneumatic InstallationReference Air Package Instrument Air Reference AirFlow SET MAX Point Knob Mounting Holes Outlet ConnectionOutlet Inlet ConnectionPower UP Section Startup and OperationGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateMenu Tree OperationOverview DisplayProgram DiagnosticsCalcheck MenuNavigation Infrared Remote Control IRCO2 Concentration O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code Program MenuCell T HI O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Upper Range VALSET O2 Filter Time Reset MAX Cell TTo Reset Maximum Cell Temperature O2 Concentration Analyzer Access CodeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 5081 Transmitter ParametersModel 5081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 5081FG Connections Hart Communicator Signal LineSection HART/AMS OverviewAnalog Output Device Method 2, For Load Resistance 250 OhmsMenu Tree for Hart Communicator TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC OFF-LINE and ON-LINE OperationsDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 5081FG Spare Board Stack Replacement Section Maintenance and ServiceModel 5081 Electronics Replacement Display Board ReplacementRibbon Cable Analog Board CPU Two-Wire In Situ Oxygen Analyzer Exploded ViewOxygen Probe Replacement Oxygen Probe Terminal Block IMPEDANCE, Ω Section TroubleshootingProbe Life Normal Operational ConditionsFault Indications OfresponseSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 5081FG Section Return of Material Model 5081FG Index No Part Number Description Replacement PartsReplacement Parts List SectionModel 5081FG Section Index Model 5081FG Model 5081FG Page Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no

5081FG specifications

The Emerson 5081FG Series is a cutting-edge digital smart transmitter designed for various industrial applications, including oil and gas, chemical processing, and water treatment. Known for its accuracy and reliability, the Emerson 5081FG integrates an array of advanced technologies to enhance measurement capabilities and operational efficiency.

One of the main features of the 5081FG is its multi-variable measurement, which allows for the integration of differential pressure, static pressure, and temperature readings into a single device. This capability significantly reduces the need for multiple instruments, leading to lower installation costs and improved data collection efficiency.

With Smart Wireless technology, the Emerson 5081FG offers robust wireless communication, making it ideal for remote or hard-to-reach locations. Its ability to transmit real-time data wirelessly enhances operational awareness and enables proactive maintenance decisions. The transmitter can easily connect to various process control systems for seamless integration and data sharing.

The 5081FG also incorporates Emerson's advanced diagnostics and predictive maintenance algorithms. These features continuously monitor the health of the transmitter and the processes it measures, alerting operators to potential issues before they escalate into serious problems. This proactive approach to maintenance not only improves safety but also minimizes downtime and maximizes productivity.

Another significant characteristic of the Emerson 5081FG is its rugged design, built to withstand harsh industrial environments. With an IP67 rating, it is resistant to dust and water ingress, ensuring reliable operation even in challenging conditions. The device is also designed to handle a wide range of temperatures and pressures, making it versatile for diverse applications.

User-friendly configuration and programming are facilitated by Emerson's intuitive software tools, which provide easy setup and calibration processes. This simplicity allows operators to optimize performance without needing extensive training or support.

In conclusion, the Emerson 5081FG is a powerful and reliable smart transmitter that addresses the needs of various industries. Its multi-variable measurement capability, wireless communication, advanced diagnostics, and rugged design make it an excellent choice for companies looking to enhance their operational efficiency and maintain high standards of safety and reliability. With the Emerson 5081FG, businesses can expect not only accurate measurements but also valuable insights that contribute to smarter decision-making.