Emerson 5081FG instruction manual SET HI Bottle O2, SET LO Bottle O2, Concentration

Page 51

Model 5081FG

Instruction Manual

IB-106-5081, Rev. 1.0 May 2005

O2 CONCENTRATION

O2 CONCENTRATION

%

%

3

 

HIGH CALIBRATION

 

CHECK GAS O2

26020015

PERCENTAGE

26020049

Figure 3-15. TRIM 20 mA?

i.TRIM 20 mA?

Refer to Figure 3-15. Use this screen to trim the 20 mA value of the 4-20 mA output.

NOTE

Before trimming the 20 mA value, you must break the loop to add the amme- ter. Power down the unit, connect the ammeter in series with Model 5081 Transmitter terminals 15(-) and 16(+), power up the unit, and return to the TRIM 4 mA? screen.

Press ENTER to begin editing. Use the ar- row keys to select and change the screen value to the value displayed on the installed ammeter. Press ENTER to accept the value. After the value is entered, the unit calibrates itself to ensure it outputs 20 mA. Both the display and the ammeter will up- date to 20 mA. Pressing EXIT returns to the initial TRIM 20 mA? screen, and pressing NEXT displays the SET HI BOTTLE O2 screen.

j.SET HI BOTTLE O2

Refer to Figure 3-16. Use this screen to identify, within the electronics, the percent- age of O2 used as the high calibration check gas. Press ENTER to begin editing. Use the arrow keys to select and change the screen value to the O2 percentage of the high cali- bration check gas. Press EN TER to accept

Figure 3-16. SET HI BOTTLE O2

the value. Press NEXT to display the SET LO BOTTLE O2 screen.

k.SET LO BOTTLE O2

Refer to Figure 3-17. Use this screen to identify, within the electronics, the percent- age of O2 used as the low calibration check gas. Press ENTER to begin editing. Use the arrow keys to select and change the screen value to the O2 percentage of the low cali- bration check gas. Press ENTER to accept the value. Press NEXT to display the SET O2 TRACKING screen.

O2 CONCENTRATION

%

LOW CALIBRATION

CHECK GAS O2

PERCENTAGE

26020050

Figure 3-17. SET LO BOTTLE O2

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation 3-9

Image 51
Contents Model 5081FG Essential Instructions Highlights of Changes Effective May, 2005 RevSummary Page Model 5081FG Table of ContentsList of Illustrations List of Tables Model 5081FG Definitions PrefaceModel 5081FG Belangrijk Vigtigt Model 5081FG Tärkeää Model 5081FG Wichtig Importante Viktig Model 5081FG Model 5081FG Viktigt Model 5081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Description System ConfigurationSystem Overview ScopeSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOperating Load Region Lift OFF Maximum Power Supply VoltageOhms VDCInspect Section InstallationPRE-INSTALLATION Mechanical InstallationSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 5081FG Installing Model 5081 Transmitter Mounting Holes Cover LockCircuit Terminal END Block TBElectrical Installation Display Positioning AssemblyModel 5081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Calibration Check Gas Pneumatic InstallationReference Air Package Instrument Air Reference AirFlow SET MAX Point Knob Mounting Holes Outlet ConnectionOutlet Inlet ConnectionPower UP Section Startup and OperationGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateMenu Tree OperationOverview DisplayProgram DiagnosticsCalcheck MenuNavigation Infrared Remote Control IRC O2 Concentration O2 Concentration Analyzer Access Code Or Security Analyzer Access Code Program MenuCell T HI O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Upper Range VALSET O2 Filter Time Reset MAX Cell TTo Reset Maximum Cell Temperature O2 Concentration Analyzer Access CodeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 5081 Transmitter ParametersModel 5081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 5081FG Connections Hart Communicator Signal LineSection HART/AMS OverviewAnalog Output Device Method 2, For Load Resistance 250 OhmsMenu Tree for Hart Communicator TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC OFF-LINE and ON-LINE OperationsDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 5081FG Spare Board Stack Replacement Section Maintenance and ServiceModel 5081 Electronics Replacement Display Board ReplacementRibbon Cable Analog Board CPU Two-Wire In Situ Oxygen Analyzer Exploded ViewOxygen Probe Replacement Oxygen Probe Terminal Block IMPEDANCE, Ω Section TroubleshootingProbe Life Normal Operational ConditionsFault Indications OfresponseSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 5081FG Section Return of Material Model 5081FG Index No Part Number Description Replacement PartsReplacement Parts List SectionModel 5081FG Section Index Model 5081FG Model 5081FG Page Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no