3.For correct viewing orientation, the dis- play may be changed 90 degrees, us- ing the following procedure:

(a)Refer to Figure 2-8. Loosen the cover lock screw until the cover lock is disengaged from the knurled surface on the threaded circuit end cap.

(b)Remove the circuit end cap.

(c)Remove the three screws retaining the display board in place.

(d)Lift and rotate the display board 90 degrees either way.

(e)Reposition the display board on the standoffs. Install and tighten all three screws.

CIRCUIT END

CAP

SCREW

DISPLAY

90

O

BOARD

 

 

 

90O

HOUSING

26020061

Figure 2-8. Display Positioning Assembly

(f) Install the circuit end cap and tighten the cover lock screw to se- cure the cover lock in place.

2-3 ELECTRICAL INSTALLATION

2

All wiring must conform to local and national

 

codes.

 

For intrinsically safe applications, re-

 

fer to drawing 1400184, page 10-2 of

 

this Instruction Bulletin.

 

Disconnect and lock out power before connecting the unit to the power supply.

Install all protective equipment covers and safety ground leads after installa- tion. Failure to install covers and ground leads could result in serious injury or death.

To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, con- nection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will discon- nect all current-carrying conductors during a fault situation. This circuit breaker should also include a me- chanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.

a.General

The power supply and signal wiring should be shielded. Also, make sure the signal wiring is grounded at the Model 5081 Transmitter end only. Do not ground the signal loop at more than one point. Twisted

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-9

Image 39
Contents Model 5081FG Essential Instructions Highlights of Changes Effective May, 2005 RevSummary Page Model 5081FG Table of ContentsList of Illustrations List of Tables Model 5081FG Definitions PrefaceModel 5081FG Belangrijk Vigtigt Model 5081FG Tärkeää Model 5081FG Wichtig Importante Viktig Model 5081FG Model 5081FG Viktigt Model 5081FG Component Checklist of Typical System Package Contents Section Description and SpecificationsCode Mounting Adapter Stack Side Product MatrixSystem Description System ConfigurationSystem Overview ScopeSystem Features Handling the Analyzer System Considerations Transmitter Typical System InstallationProbe SpecificationsOperating Load Region Lift OFF Maximum Power Supply VoltageOhms VDCInspect Section InstallationPRE-INSTALLATION Mechanical InstallationSide View Bottom View Front ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 5081FG Installing Model 5081 Transmitter Mounting Holes Cover LockCircuit Terminal END Block TBElectrical Installation Display Positioning AssemblyModel 5081 Transmitter 4-20 mA and Signal Connections Terminal Block ConduitThermocouple CELL-WH CELL+BK Calibration Check Gas Pneumatic InstallationReference Air Package Instrument Air Reference AirFlow SET MAX Point Knob Mounting Holes Outlet ConnectionOutlet Inlet ConnectionPower UP Section Startup and OperationGeneral Proper Calibration Check Gas Flow Rate Reestablishing Proper Calibration Check GAS Flow RateMenu Tree OperationOverview DisplayProgram DiagnosticsCalcheck MenuNavigation Infrared Remote Control IRCO2 Concentration O2 Concentration Analyzer Access Code Or SecurityAnalyzer Access Code Program MenuCell T HI O2 Concentration Indicator Fault ValueO2 Concentration Upper Cell Temperature Fault Value Upper Range VALSET O2 Filter Time Reset MAX Cell TTo Reset Maximum Cell Temperature O2 Concentration Analyzer Access CodeSET HI Bottle O2 SET LO Bottle O2Concentration Parameter Range Default Model 5081 Transmitter ParametersModel 5081 Transmitter Parameters SET Code Diagnostics Menu Show FaultC mV 24. Previous Slope Previous Constant 23. Cell ImpedanceCalcheck Menu Accept LOW O2 27. in MANUAL?29. Accept LOW O2 Slope Model 5081FG Connections Hart Communicator Signal LineSection HART/AMS OverviewAnalog Output Device Method 2, For Load Resistance 250 OhmsMenu Tree for Hart Communicator TWO-WIRE in Situ Oxygen AnalyzerHart Communicator PC OFF-LINE and ON-LINE OperationsDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 5081FG Spare Board Stack Replacement Section Maintenance and ServiceModel 5081 Electronics Replacement Display Board ReplacementRibbon Cable Analog Board CPU Two-Wire In Situ Oxygen Analyzer Exploded ViewOxygen Probe Replacement Oxygen Probe Terminal Block IMPEDANCE, Ω Section TroubleshootingProbe Life Normal Operational ConditionsFault Indications OfresponseSpeed Thermocouple Identifying and Correcting Fault IndicationsFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 6, Cell Impedance Too High Fault 5, O2 Cell Open11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 CellModel 5081FG Section Return of Material Model 5081FG Index No Part Number Description Replacement PartsReplacement Parts List SectionModel 5081FG Section Index Model 5081FG Model 5081FG Page Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no

5081FG specifications

The Emerson 5081FG Series is a cutting-edge digital smart transmitter designed for various industrial applications, including oil and gas, chemical processing, and water treatment. Known for its accuracy and reliability, the Emerson 5081FG integrates an array of advanced technologies to enhance measurement capabilities and operational efficiency.

One of the main features of the 5081FG is its multi-variable measurement, which allows for the integration of differential pressure, static pressure, and temperature readings into a single device. This capability significantly reduces the need for multiple instruments, leading to lower installation costs and improved data collection efficiency.

With Smart Wireless technology, the Emerson 5081FG offers robust wireless communication, making it ideal for remote or hard-to-reach locations. Its ability to transmit real-time data wirelessly enhances operational awareness and enables proactive maintenance decisions. The transmitter can easily connect to various process control systems for seamless integration and data sharing.

The 5081FG also incorporates Emerson's advanced diagnostics and predictive maintenance algorithms. These features continuously monitor the health of the transmitter and the processes it measures, alerting operators to potential issues before they escalate into serious problems. This proactive approach to maintenance not only improves safety but also minimizes downtime and maximizes productivity.

Another significant characteristic of the Emerson 5081FG is its rugged design, built to withstand harsh industrial environments. With an IP67 rating, it is resistant to dust and water ingress, ensuring reliable operation even in challenging conditions. The device is also designed to handle a wide range of temperatures and pressures, making it versatile for diverse applications.

User-friendly configuration and programming are facilitated by Emerson's intuitive software tools, which provide easy setup and calibration processes. This simplicity allows operators to optimize performance without needing extensive training or support.

In conclusion, the Emerson 5081FG is a powerful and reliable smart transmitter that addresses the needs of various industries. Its multi-variable measurement capability, wireless communication, advanced diagnostics, and rugged design make it an excellent choice for companies looking to enhance their operational efficiency and maintain high standards of safety and reliability. With the Emerson 5081FG, businesses can expect not only accurate measurements but also valuable insights that contribute to smarter decision-making.