Emerson 5081FG instruction manual in MANUAL?, Accept LOW O2

Page 56

Instruction Manual

IB-106-5081, Rev. 1.0 May 2005

Model 5081FG

a.IN MANUAL? Refer to Figure 3-27.

Failure to remove the analyzer from automatic control loops prior to per- forming this procedure may result in a dangerous operating condition.

If the O2 output value is used in any auto- matic process control loops, the loop must be placed in manual before beginning a calibration check.

O2 CONCENTRATION

%

%

HIGH CALIBRATION

CHECK GAS O2 VALUE

Once the analyzer is removed from any automatic control loops, press ENTER to edit the screen. Use the arrow keys to se- lect Y (yes) and press ENTER to process the selection and to display the ACCEPT HIGH O2 screen.

O2 CONCENTRATION

%

Y/N (FOR USER

RESPONSE)

26020054

Figure 3-27. IN MANUAL?

26020055

Figure 3-28. ACCEPT HIGH O2

b.ACCEPT HIGH O2

Refer to Figure 3-28. After pressing ENTER to begin the calibration check, the high cali- bration check gas starts to flow. After wait- ing approximately three minutes for the displayed O2 value to settle, press NEXT to accept the high calibration check gas read- ing and apply the low calibration check gas. The next screen to display is the ACCEPT LOW O2 screen.

c.ACCEPT LOW O2

Refer to Figure 3-29.Once the low calibra- tion check gas is applied, wait approxi- mately three minutes for the displayed O2 value to settle. Once the value settles, press NEXT to accept the reading and to display the SLOPE screen.

3-14 Startup and Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 56
Contents Model 5081FG Essential Instructions Summary Highlights of ChangesEffective May, 2005 Rev Page Table of Contents Model 5081FGList of Illustrations List of Tables Model 5081FG Preface DefinitionsModel 5081FG Belangrijk Vigtigt Model 5081FG Tärkeää Model 5081FG Wichtig Importante Viktig Model 5081FG Model 5081FG Viktigt Model 5081FG Section Description and Specifications Component Checklist of Typical System Package ContentsProduct Matrix Code Mounting Adapter Stack SideSystem Configuration System OverviewScope System DescriptionSystem Features Handling the Analyzer System Considerations Typical System Installation TransmitterSpecifications ProbeLift OFF Maximum Power Supply Voltage OhmsVDC Operating Load RegionSection Installation PRE-INSTALLATIONMechanical Installation InspectFront View Side View Bottom ViewMetal Wall Masonry Wall Stack or Duct Horizontal Probe Installation Adjusting Probe Insertion Depth Model 5081FG Installing Model 5081 Transmitter Cover Lock Circuit Terminal ENDBlock TB Mounting HolesDisplay Positioning Assembly Electrical InstallationThermocouple CELL-WH CELL+BK Model 5081 Transmitter 4-20 mA and Signal ConnectionsTerminal Block Conduit Pneumatic Installation Reference Air PackageInstrument Air Reference Air Calibration Check GasOutlet Connection OutletInlet Connection Flow SET MAX Point Knob Mounting HolesGeneral Power UPSection Startup and Operation Reestablishing Proper Calibration Check GAS Flow Rate Proper Calibration Check Gas Flow RateOperation OverviewDisplay Menu TreeDiagnostics CalcheckMenu ProgramInfrared Remote Control IRC NavigationO2 Concentration Analyzer Access Code Or Security Analyzer Access CodeProgram Menu O2 ConcentrationO2 Concentration Indicator Fault Value O2 Concentration Upper Cell Temperature Fault ValueUpper Range VAL Cell T HIReset MAX Cell T To Reset Maximum Cell TemperatureO2 Concentration Analyzer Access Code SET O2 Filter TimeConcentration SET HI Bottle O2SET LO Bottle O2 Model 5081 Transmitter Parameters SET Code Parameter Range DefaultModel 5081 Transmitter Parameters C mV Diagnostics MenuShow Fault 23. Cell Impedance 24. Previous Slope Previous ConstantCalcheck Menu 27. in MANUAL? Accept LOW O229. Accept LOW O2 Slope Model 5081FG Hart Communicator Signal Line Section HART/AMSOverview ConnectionsMethod 2, For Load Resistance 250 Ohms Analog Output DeviceTWO-WIRE in Situ Oxygen Analyzer Hart Communicator PCOFF-LINE and ON-LINE Operations Menu Tree for Hart CommunicatorDevice Setup From Sheet Basic Setup Detailed From Sheet Signal Condition Detailed Setup Output Hart Communicator Start CAL- Check Method Model 5081FG Section Maintenance and Service Model 5081 Electronics ReplacementDisplay Board Replacement Spare Board Stack ReplacementTwo-Wire In Situ Oxygen Analyzer Exploded View Ribbon Cable Analog Board CPUOxygen Probe Replacement Oxygen Probe Terminal Block Section Troubleshooting Probe LifeNormal Operational Conditions IMPEDANCE, ΩSpeed Fault IndicationsOfresponse Identifying and Correcting Fault Indications ThermocoupleFault 1, Open Thermocouple Fault 3, Shorted Thermocouple Fault 5, O2 Cell Open Fault 5, O2 Cell Open Fault 6, Cell Impedance Too HighFault 7, Reversed O2 Cell 11. Fault 7, Reversed O2 CellModel 5081FG Section Return of Material Model 5081FG Replacement Parts Replacement Parts ListSection Index No Part Number DescriptionModel 5081FG Section Index Model 5081FG Model 5081FG Page Warranty Two-Wire In Situ Oxygen Analyzer Serial no Order no

5081FG specifications

The Emerson 5081FG Series is a cutting-edge digital smart transmitter designed for various industrial applications, including oil and gas, chemical processing, and water treatment. Known for its accuracy and reliability, the Emerson 5081FG integrates an array of advanced technologies to enhance measurement capabilities and operational efficiency.

One of the main features of the 5081FG is its multi-variable measurement, which allows for the integration of differential pressure, static pressure, and temperature readings into a single device. This capability significantly reduces the need for multiple instruments, leading to lower installation costs and improved data collection efficiency.

With Smart Wireless technology, the Emerson 5081FG offers robust wireless communication, making it ideal for remote or hard-to-reach locations. Its ability to transmit real-time data wirelessly enhances operational awareness and enables proactive maintenance decisions. The transmitter can easily connect to various process control systems for seamless integration and data sharing.

The 5081FG also incorporates Emerson's advanced diagnostics and predictive maintenance algorithms. These features continuously monitor the health of the transmitter and the processes it measures, alerting operators to potential issues before they escalate into serious problems. This proactive approach to maintenance not only improves safety but also minimizes downtime and maximizes productivity.

Another significant characteristic of the Emerson 5081FG is its rugged design, built to withstand harsh industrial environments. With an IP67 rating, it is resistant to dust and water ingress, ensuring reliable operation even in challenging conditions. The device is also designed to handle a wide range of temperatures and pressures, making it versatile for diverse applications.

User-friendly configuration and programming are facilitated by Emerson's intuitive software tools, which provide easy setup and calibration processes. This simplicity allows operators to optimize performance without needing extensive training or support.

In conclusion, the Emerson 5081FG is a powerful and reliable smart transmitter that addresses the needs of various industries. Its multi-variable measurement capability, wireless communication, advanced diagnostics, and rugged design make it an excellent choice for companies looking to enhance their operational efficiency and maintain high standards of safety and reliability. With the Emerson 5081FG, businesses can expect not only accurate measurements but also valuable insights that contribute to smarter decision-making.