Drive Medical Design 18600 user manual Maintenance10, Troubleshooting, Problem Action

Page 10

maintenance10

!NOTE

Use only Drive filters. Do not substitute any other material for this bacteria filter. Improper substitution may lead to contamination or poor performance;

Fuse Change:

1.Use only the same type of fuse to replace the broken fuse.

2.Fuse specification: Miniature Glass Fuse 5x20mm, Fast Blow (Quick Acting), F2A/250V.

!WARNING

When this unit or accessories reach the end of their useful lives (depends on using conditions), residues in collection bottle, tubing and filter may cause respiratory infection. Dump the wasting collection bottle, tubing and filter into a specific waste bin labeled with medical waste.

troubleshooting

Your Heavy Duty Suction Pump contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO THE MEDICAL EQUIPMENT PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:

PROBLEM

ACTION

Unit does not turn on

1. Check power source and connection

 

2. Ensure that the wall outlet is live

 

3. Check power cord is not damage

Motor runs, no vacuum

1. Verify tubing connection security

 

2. Check for leaks or tubing kinks

 

3. Ensure that float shut-off is not activated

 

4. Check for bottle leaks and cracks

Low Vacuum

1. Use vacuum adjust knob to increase vacuum level

 

2. Check system for leaks

 

3. Adjust vacuum adjust knob and release

Image 10
Contents USER’S Manual Manual DEL Usuario Guide DE L’UTILISATEUR How to Operate Your Heavy Duty Suction Pump IndexSafety information Safety InformationSpecific Warnings AC Power SourceElectrical Introduction Illustration of important parts Set up & operation HOW to Operate Your Heavy Duty Suction PumpCleaning Cleaning & maintenance MaintenanceTroubleshooting Maintenance10Problem Action Specifications/classifications Warranty12 Two Years Limited WarrantyPrecaución IndiceNota Peligro Información de seguridadInformación DE Seguridad AdvertenciaAdvertencias Específicas Fuente DE Alimentación DE CA PrecauciónIntroducción Diagramas de partes importantes Configuración & operación Cómo Operar SU Bomba DE Succión DE USO PesadoLimpieza Limpieza & mantenimiento MantenimientoMantenimiento21 Solución de problemasProblema AcciónEspecificaciones/clasificaciones Garantía23 Dos Años de Garantía LimitadaTable des matières Mise EN GardeRenseignements Concernant LA Sécurité AvertissementAvertissements Spécifiques Alimentation Électrique CA ! Mise EN GardeIntroduction Diagramme des principales pièces Installation et fonctionnement Utilisation DE Votre Pompe Aspiratrice RobusteInstallation et fonctionnement Nettoyage et entretien EntretienDépannage SolutionEntretien32 ProblèmeSpécifications / classification Garantie34 Garantie Limitée de Deux AnsPage Seaview Boulevard Port Washington NY Phone Fax