set up & operation | 8 |
|
4.Place the patient tubing to the appropriate position of patient and start to suck sputum.
5.When the sputum reaches the safe full level of collection bottle (800 cc), the suctioning will stop sucking automatically (float
6.Switch the power switch to “O” position (off), shut off the suction motor.
7. Remove the sputum from collection bottle after the motor stops turning completely. (Please refer to the section “CLEAN INSTRUCTIONS” on page 9 for the directions of how to clean the collection bottle).
8. Turn on the unit and start to run the suction unit again.
!NOTE
1.If the unit sustains a severe drop, validity of the gauge must be checked. Inspect if the float
2.Always transport unit with vacuum adjust knob rotated fully clockwise in case unit is dropped.
!CAUTION
1.When automatic float
2.Should fluid be aspirated back into the unit, send your suction to an authorized service center for a detailed inspection as the vacuum pump may have been damaged.
cleaning
Clean Collection Bottle:
1.Turn off the suction unit and allow vacuum to drop, then disconnect AC power source.
2.With the collection bottle still in the holder, remove the collection bottle lid. The bottle can now be taken out of the holder to be emptied.
!NOTE
Collection bottle should be emptied after each use.
3.Collection bottle should be thoroughly cleaned after each use by one of the following three methods:
a. Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry.
b. Wash with rubbing alcohol and air dry
c. Wash with a commercial