Drive Medical Design 18600 user manual Cleaning

Page 8

set up & operation

8

 

4.Place the patient tubing to the appropriate position of patient and start to suck sputum.

5.When the sputum reaches the safe full level of collection bottle (800 cc), the suctioning will stop sucking automatically (float shut-off).

6.Switch the power switch to “O” position (off), shut off the suction motor.

7. Remove the sputum from collection bottle after the motor stops turning completely. (Please refer to the section “CLEAN INSTRUCTIONS” on page 9 for the directions of how to clean the collection bottle).

8. Turn on the unit and start to run the suction unit again.

!NOTE

1.If the unit sustains a severe drop, validity of the gauge must be checked. Inspect if the float shut-off is still functioning properly when the liquid in the collection bottle reaches the safe full level.

2.Always transport unit with vacuum adjust knob rotated fully clockwise in case unit is dropped.

!CAUTION

1.When automatic float shut-off is activated, contents of the collection bottles should be emp- tied. Further suctioning could cause damage to the vacuum pump.

2.Should fluid be aspirated back into the unit, send your suction to an authorized service center for a detailed inspection as the vacuum pump may have been damaged.

cleaning

Clean Collection Bottle:

1.Turn off the suction unit and allow vacuum to drop, then disconnect AC power source.

2.With the collection bottle still in the holder, remove the collection bottle lid. The bottle can now be taken out of the holder to be emptied.

!NOTE

Collection bottle should be emptied after each use.

3.Collection bottle should be thoroughly cleaned after each use by one of the following three methods:

a. Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry.

b. Wash with rubbing alcohol and air dry

c. Wash with a commercial (bacterial-germicidal) disinfectant, follow disinfectant manufacturer’s recommended instructions and dilution rates carefully.

Image 8
Contents USER’S Manual Manual DEL Usuario Guide DE L’UTILISATEUR How to Operate Your Heavy Duty Suction Pump IndexSafety information Safety InformationElectrical AC Power SourceSpecific Warnings Introduction Illustration of important parts Set up & operation HOW to Operate Your Heavy Duty Suction PumpCleaning Cleaning & maintenance MaintenanceProblem Action Maintenance10Troubleshooting Specifications/classifications Warranty12 Two Years Limited WarrantyNota IndicePrecaución Información de seguridad Información DE SeguridadPeligro AdvertenciaAdvertencias Específicas Fuente DE Alimentación DE CA PrecauciónIntroducción Diagramas de partes importantes Configuración & operación Cómo Operar SU Bomba DE Succión DE USO PesadoLimpieza Limpieza & mantenimiento MantenimientoSolución de problemas ProblemaMantenimiento21 AcciónEspecificaciones/clasificaciones Garantía23 Dos Años de Garantía LimitadaTable des matières Mise EN GardeRenseignements Concernant LA Sécurité AvertissementAvertissements Spécifiques Alimentation Électrique CA ! Mise EN GardeIntroduction Diagramme des principales pièces Installation et fonctionnement Utilisation DE Votre Pompe Aspiratrice RobusteInstallation et fonctionnement Nettoyage et entretien EntretienSolution Entretien32Dépannage ProblèmeSpécifications / classification Garantie34 Garantie Limitée de Deux AnsPage Seaview Boulevard Port Washington NY Phone Fax