Drive Medical Design 18600 user manual Cleaning & maintenance, Maintenance

Page 9

cleaning & maintenance

9

 

Clean Suction Unit and Collection Bottle Lid:

1.Disconnect the suction unit from all external power sources.

2.Wipe the suction unit housing and collection bottle lid with a clean cloth and any commer- cial (bacterial-germicidal) disinfectant.

!CAUTION

Do not submerge in water as this will result in severe damage to the vacuum pump.

Clean Tubing:

1.Disconnect the tubing from the unit.

2.Tubing should be rinsed thoroughly after every use by running hot tap water through it followed by a solution of one part vinegar to three parts hot water.

3.Rinse with hot tap water and air dry.

4.Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.

!WARNING

The waste in collection bottle may cause respiratory infection which must be handled according to medical waste disposing regulation. Operator must wear gloves and medical class mask be- fore cleaning the collection bottle, tubing and handle the wasting unit and/or accessories.

MAINTENANCE

Inspect suction tubing and collection bottle for leaks, cracks, etc. before each use.

!DANGER

Electric shock hazard

Do not attempt to open or remove the suction cabinet. If a service is required, return your suction to an authorized service center. Opening or tampering with the unit will void warranty.

Bacteria Filter Change:

1.Bacteria filter should be replaced as needed. If overflow occurs, change the filter immediately.

2.Remove filter by disconnecting it from tubing connected to pressure adjusting valve and tubing connected to collection bottle lid assembly.

3.Replace with a new Drive bacteria filter and connect to above two tubing. Additional filters can be purchased from an authorized dealer.

Image 9
Contents USER’S Manual Manual DEL Usuario Guide DE L’UTILISATEUR Index How to Operate Your Heavy Duty Suction PumpSafety Information Safety informationAC Power Source Specific WarningsElectrical Introduction Illustration of important parts HOW to Operate Your Heavy Duty Suction Pump Set up & operationCleaning Maintenance Cleaning & maintenanceMaintenance10 TroubleshootingProblem Action Specifications/classifications Two Years Limited Warranty Warranty12Indice PrecauciónNota Información DE Seguridad Información de seguridadPeligro AdvertenciaFuente DE Alimentación DE CA Precaución Advertencias EspecíficasIntroducción Diagramas de partes importantes Cómo Operar SU Bomba DE Succión DE USO Pesado Configuración & operaciónLimpieza Mantenimiento Limpieza & mantenimientoProblema Solución de problemasMantenimiento21 AcciónEspecificaciones/clasificaciones Dos Años de Garantía Limitada Garantía23Mise EN Garde Table des matièresAvertissement Renseignements Concernant LA SécuritéAlimentation Électrique CA ! Mise EN Garde Avertissements SpécifiquesIntroduction Diagramme des principales pièces Utilisation DE Votre Pompe Aspiratrice Robuste Installation et fonctionnementInstallation et fonctionnement Entretien Nettoyage et entretienEntretien32 SolutionDépannage ProblèmeSpécifications / classification Garantie Limitée de Deux Ans Garantie34Page Seaview Boulevard Port Washington NY Phone Fax