Drive Medical Design 18600 Set up & operation, HOW to Operate Your Heavy Duty Suction Pump

Page 7

set up & operation

7

 

1.Connect either end of the connecting tubing (item# 9) to the top of collection bottle connector then connect the other end to the bacteria filter.

2.The patient tubing should be connected to collection bottle at the outlet labeled (patient). Please assure that all connections are secure and without leaks before using.

7. HOW TO OPERATE YOUR HEAVY DUTY SUCTION PUMP

1.Before use, check the specification label on the side of the suction unit to ensure that the volt- age and current indicated on the unit correspond with the voltage and current available. Make sure the power switch located on the side of the unit is in the “Off” position before connecting the unit to a power source.

2.Switch the power switch to “I” position (on), the unit start running.

1. Off position

2. On position

3.Adjust the vacuum level from 0 to 560 mm Hg by turning the vacuum adjust knob (clockwise to increase vacuum and counter-clockwise to decrease vacuum). Referring to the vacuum gauge while setting the desired level of vacuum. To accurately read the gauge, block the patient end of the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum reading.

Image 7 Drive Medical Design 18600 user manual Set up & operation, HOW to Operate Your Heavy Duty Suction Pump
Contents USER’S Manual Manual DEL Usuario Guide DE L’UTILISATEUR Index How to Operate Your Heavy Duty Suction PumpSafety Information Safety informationSpecific Warnings AC Power SourceElectrical Introduction Illustration of important parts HOW to Operate Your Heavy Duty Suction Pump Set up & operationCleaning Maintenance Cleaning & maintenanceTroubleshooting Maintenance10Problem Action Specifications/classifications Two Years Limited Warranty Warranty12Precaución IndiceNota Advertencia Información de seguridadInformación DE Seguridad PeligroFuente DE Alimentación DE CA Precaución Advertencias EspecíficasIntroducción Diagramas de partes importantes Cómo Operar SU Bomba DE Succión DE USO Pesado Configuración & operaciónLimpieza Mantenimiento Limpieza & mantenimientoAcción Solución de problemasProblema Mantenimiento21Especificaciones/clasificaciones Dos Años de Garantía Limitada Garantía23Mise EN Garde Table des matièresAvertissement Renseignements Concernant LA SécuritéAlimentation Électrique CA ! Mise EN Garde Avertissements SpécifiquesIntroduction Diagramme des principales pièces Utilisation DE Votre Pompe Aspiratrice Robuste Installation et fonctionnementInstallation et fonctionnement Entretien Nettoyage et entretienProblème SolutionEntretien32 DépannageSpécifications / classification Garantie Limitée de Deux Ans Garantie34Page Seaview Boulevard Port Washington NY Phone Fax