Altek 830-KP manual Calibrate RTD Inputs, Two Wire RTD Connection, Three Wire RTD Connection

Page 10

OPERATING INSTRUCTIONS

CALIBRATE RTD INPUTS

Choose this function to simulate a temperature signal into any instrument requiring an RTD input.

1)Disconnect the RTD from the instrument being calibrated.

2)Press the T/C/RTD/pushbutton until any RTD type (Pt, Ni or Cu) appears on the display.

3)Press the DISPLAY/SOURCE/READ push-button until SOURCE and any RTD TYPE appear on the display

4)Connect using 2 or 3 wires as in the diagrams on the opposite page. Spade lugs are recommended to minimize any contact resistance.

Output temperature is continuously adjustable with the UP/DOWN push-buttons. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs”.

To Change the RTD type

1)Press the TYPE/ENG UNITS push-button. The words SELECT RTD TYPE and the current selection appear on the display.

2)Press any UP/DOWN push-button until the required type appears on the display.

3)Press the STORE/RESET push-button or wait 5 seconds to store the selection

Note on 4-Wire Connections:

The TechChek 830-KP may be used to calibrate instruments requiring a 4-Wire RTD connection. This connection is valid only when the lead wires included with the 830-KP are used to connect the 830-KP directly to the instrument. Use of longer wires can cause errors in the output setting.

For long 4-Wire cable runs use the 3-Wire connection and add a stacking banana jack (not available from Transmation) to connect a second wire to jack #1. Connect this extra wire to the fourth wire field connection.

OPERATING INSTRUCTIONS

CALIBRATE RTD INPUTS

Two Wire RTD Connection

Receiver Input

Ohms or 2-Wire RTD

Controller

Transmitter

Computer

Three Wire RTD Connection

Instrument with

RTD Input

Controller

Temperature Transmitter

Temperature Indicator

Temperature Trip or Alarm

Four Wire RTD Connection

Jumper

+ COMP

}SENSOR

- COMP

Jumper

16

17

Image 10
Contents Process Calibrator Contents General InformationGeneral TURN-ON Rear Label Condensed Operation GuideConnections Operating Instructions Display Backlighting PowerGeneral Display Contrast Turn OFFRecalling QUIK-CHEK Outputs Source ModeStoring QUIK-CHEK Outputs Calibrate Milliamp InputsWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Simulate 2-WIRE TransmittersCalibrate Voltage Inputs Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Thermocouple Inputs Calibrate Resistance InputsFour Wire RTD Connection Two Wire RTD ConnectionThree Wire RTD Connection Calibrate RTD InputsRead Functions OUT of Range SignalsCalibrate Frequency Inputs MIN/MAXRead DC and AC Voltage Outputs MV, V, Vhi and Vac Read Milliamp OutputsConnection for DC Volts to 200.0 vhi Read AC Voltages Measure Thermocouple SensorsOhms Read ResistanceCheck Continuity Black Measure RTD SensorsConnection for signals with amplitudes to 250 V RMS Count FrequenciesCalibrate Transmitters ANY Source Function Calibrate 2-WIRE TransmittersCalibrate 4-WIRE Transmitters Read mA, Read %, Read DP%Read Presssure Fitting InstallationRead Pressure Calibrate Pressure Transmitters Read Pressure Specifications Read Thermocouples Measure DC Volts RangesCold Junction Accuracy ±1C Read OhmsThermocouple Specifications Type Range AccuracyRTD Specifications RTD Type Alpha Range AccuracyWarranty Ordering InformationPN 1800470 Rev a April