OPERATING INSTRUCTIONS
CALIBRATE RTD INPUTS
Choose this function to simulate a temperature signal into any instrument requiring an RTD input.
1)Disconnect the RTD from the instrument being calibrated.
2)Press the T/C/RTD/Ω pushbutton until any RTD type (Pt, Ni or Cu) appears on the display.
3)Press the DISPLAY/SOURCE/READ
4)Connect using 2 or 3 wires as in the diagrams on the opposite page. Spade lugs are recommended to minimize any contact resistance.
Output temperature is continuously adjustable with the UP/DOWN
To Change the RTD type
1)Press the TYPE/ENG UNITS
2)Press any UP/DOWN
3)Press the STORE/RESET
Note on
The TechChek
For long
OPERATING INSTRUCTIONS
CALIBRATE RTD INPUTS
Two Wire RTD Connection
Receiver Input
Ohms or
Controller
Transmitter
Computer
Three Wire RTD Connection
Instrument with
RTD Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Four Wire RTD Connection
Jumper
+ COMP
}SENSOR
- COMP
Jumper
16 | 17 |