Altek 830-KP General Display Contrast, Display Backlighting, Turn OFF, Power, Selecting Functions

Page 5

OPERATING INSTRUCTIONS

GENERAL

DISPLAY CONTRAST

The contrast of the Liquid Crystal Display may adjusted for best readability.

1)Press and hold the DISPLAY/SOURCE/READ push-button while turning the 830-KP on to adjust the contrast of the display.

2)Press any UP or DOWN key until the display is most legible.

3)Press the STORE/RESET push-button or wait five seconds for your selection to be stored.

DISPLAY BACKLIGHTING

The LCD may have too little contrast where the lighting is dim or the 830-KP is in shadow. Turn on the LCD backlighting to make the display easy to read.

1)If the TechChek 830-KP is off, turn it on and wait for the display to come up in normal mode.

2)Press and hold the DISPLAY/SOURCE/READ push-button for three seconds and the backlight will turn on.

To extend the life of the batteries, it is recommended that the backlighting be turned off when the LCD is in normal light. Press and hold the DISPLAY/SOURCE/READ push-button for three seconds and the backlight will turn off.

 

TURN OFF

 

POWER

Press the POWER push-button to turn the

 

830-KP off. If AUTO-OFF is enabled, the

 

830-KP will turn itself off after 30 minutes of

 

 

inactivity.

OPERATING INSTRUCTIONS

 

 

 

GENERAL

SELECTING FUNCTIONS

 

 

 

 

Press a function key to choose among mA,

 

 

 

V, T/C, , RTD, FREQ and PRESSURE.

SELECTING RANGES

 

 

TYPE

Press the RANGE/TYPE pushbutton to

 

 

ENG UNITS

select the desired range and scale.

SOURCE RANGES

Milliamp:

Source: mA, %mA (% 4-20), DP% (% Differential Pressure)

2-Wire Simulator: mA, %mA(% 4-20), DP% (% Differential Pressure)

VDC: mV, V

T/C: Types J, T, E, K, N, R & S in °C &°F

Ohms: Ohms

RTD: Four Pt 100, Ni 120& Cu 10in °C &°F

Frequency: KHz, Hz, CPM

READ RANGES

Milliamp: mA, %mA (% 4-20), DP% (% Differential Pressure)

VDC: mV, 10V, 200V

T/C: Types J, T, E, K, N, R & S in °C &°F

Ohms: Ohms

RTD: Four Pt 100, Ni 120& Cu 10in °C &°F

Frequency: KHz, Hz, CPM

AC Volts: VAC

TRANSMITTER RANGES

Pwr Xmtr: Supplies nominal 24 VDC to power the transmitter while measuring the transmitter milliamp signal . Can be simul- taneously displayed with any Source Range or Pressure Measurement

Read mA: Measures transmitter output

6

7

Image 5
Contents Process Calibrator General Information ContentsConnections Rear Label Condensed Operation GuideGeneral TURN-ON Operating Instructions General Display Contrast PowerDisplay Backlighting Turn OFFStoring QUIK-CHEK Outputs Source ModeRecalling QUIK-CHEK Outputs Calibrate Milliamp InputsCalibrate Voltage Inputs Simulate 2-WIRE TransmittersWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Resistance Inputs Calibrate Thermocouple InputsThree Wire RTD Connection Two Wire RTD ConnectionFour Wire RTD Connection Calibrate RTD InputsCalibrate Frequency Inputs OUT of Range SignalsRead Functions MIN/MAXConnection for DC Volts to 200.0 vhi Read Milliamp OutputsRead DC and AC Voltage Outputs MV, V, Vhi and Vac Measure Thermocouple Sensors Read AC VoltagesCheck Continuity Read ResistanceOhms Measure RTD Sensors BlackCount Frequencies Connection for signals with amplitudes to 250 V RMSCalibrate 4-WIRE Transmitters Calibrate 2-WIRE TransmittersCalibrate Transmitters ANY Source Function Read mA, Read %, Read DP%Read Pressure Fitting InstallationRead Presssure Calibrate Pressure Transmitters Read Pressure Specifications Cold Junction Accuracy ±1C Measure DC Volts RangesRead Thermocouples Read OhmsType Range Accuracy Thermocouple SpecificationsRTD Type Alpha Range Accuracy RTD SpecificationsOrdering Information WarrantyPN 1800470 Rev a April