OPERATING INSTRUCTIONS
READ RESISTANCE
Ohms
Choose this function to measure resistance from 0.0 to 1000.0 Ohms.
1)Disconnect one or both input wires from the device to be calibrated.
2)Press the T/C/RTD/Ω pushbutton until the Ω symbol appears on the display.
3)Press the DISPLAY/SOURCE/READ
4)Press the TYPE/ENG UNITS
5)Connect the red READ (+) lead and the black READ
Signals above or below those available for the currently select- ed range will be indicated by OVER and UNDER on the display.
Receiver Input
Ohms or
Controller
Transmitter
Computer
OPERATING INSTRUCTIONS
CHECK CONTINUITY
Choose this function to check continuity. A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms.
1)Plug the leads into the TechChek
2)Press the T/C/RTD/Ω pushbutton until the Ω symbol appears on the display.
3)Press the DISPLAY/SOURCE/READ
4)Press the TYPE/ENG UNITS pushbuuton once to switch between READ and CONTINUITY on the display
24 | 25 |