OPERATING INSTRUCTIONS

READ RESISTANCE

Ohms

Choose this function to measure resistance from 0.0 to 1000.0 Ohms.

1)Disconnect one or both input wires from the device to be calibrated.

2)Press the T/C/RTD/pushbutton until the symbol appears on the display.

3)Press the DISPLAY/SOURCE/READ push-button until READ or CONTINUITY appears on the display

4)Press the TYPE/ENG UNITS push-button once to switch between READ and CONTINUITY on the display

5)Connect the red READ (+) lead and the black READ (-) lead of the calibrator across the resistance to be measured.

Signals above or below those available for the currently select- ed range will be indicated by OVER and UNDER on the display.

Receiver Input

Ohms or 2-Wire RTD

Controller

Transmitter

Computer

OPERATING INSTRUCTIONS

CHECK CONTINUITY

Choose this function to check continuity. A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms.

1)Plug the leads into the TechChek 830-KP as shown below.

2)Press the T/C/RTD/pushbutton until the symbol appears on the display.

3)Press the DISPLAY/SOURCE/READ push-button until READ or CONTINUITY appears on the display

4)Press the TYPE/ENG UNITS pushbuuton once to switch between READ and CONTINUITY on the display

24

25

Image 14
Contents Process Calibrator Contents General InformationConnections Rear Label Condensed Operation GuideGeneral TURN-ON Operating Instructions Display Backlighting PowerGeneral Display Contrast Turn OFFRecalling QUIK-CHEK Outputs Source ModeStoring QUIK-CHEK Outputs Calibrate Milliamp InputsCalibrate Voltage Inputs Simulate 2-WIRE TransmittersWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Thermocouple Inputs Calibrate Resistance InputsFour Wire RTD Connection Two Wire RTD ConnectionThree Wire RTD Connection Calibrate RTD InputsRead Functions OUT of Range SignalsCalibrate Frequency Inputs MIN/MAXConnection for DC Volts to 200.0 vhi Read Milliamp OutputsRead DC and AC Voltage Outputs MV, V, Vhi and Vac Read AC Voltages Measure Thermocouple SensorsCheck Continuity Read ResistanceOhms Black Measure RTD SensorsConnection for signals with amplitudes to 250 V RMS Count FrequenciesCalibrate Transmitters ANY Source Function Calibrate 2-WIRE TransmittersCalibrate 4-WIRE Transmitters Read mA, Read %, Read DP%Read Pressure Fitting InstallationRead Presssure Calibrate Pressure Transmitters Read Pressure Specifications Read Thermocouples Measure DC Volts RangesCold Junction Accuracy ±1C Read OhmsThermocouple Specifications Type Range AccuracyRTD Specifications RTD Type Alpha Range AccuracyWarranty Ordering InformationPN 1800470 Rev a April