Altek 830-KP manual Calibrate Transmitters ANY Source Function, Read mA, Read %, Read DP%

Page 17

OPERATING INSTRUCTIONS

CALIBRATE TRANSMITTERS

ANY SOURCE FUNCTION AND

READ mA, READ %, READ DP%

Choose this function to supply the input signal to the transmitter and displaying the 4-20 mA output of the transmitter (used with 4- Wire Transmitters).

ANY SOURCE FUNCTION AND

P-XMTR mA, P-XMTR %, P-XMTR DP%

Choose this function to simultaneously supply power to a 2-Wire transmitter while supply the input signal to the transmitter and display the 4-20 mA output of the transmitter.

1)Disconnect both the input and output wires from the 2-Wire Transmitter to be calibrated

2)Choose a function of the 830-KP that matches the input of the transmitter (T/C, RTD, Freq, etc.). Select the proper type, range and temperature scale if applicable.

3)Press the DISPLAY/SOURCE/READ push-button until READ or P-XMTR appears in the lower half of the display

4)Press READ/POWER TRANSMITTER to switch between reading milliamps and supply voltage to power the transmitter

5)Connect the output of the 830-KP to the signal input of the trans- mitter (using the 1, 2 & 4 connectors or the T/C connector on the 830-KP)

6)Connect the red POWER lead of the 830-KP (Connector 5) to the plus (+) output of the tranmitter and the black POWER lead

(Connector 6) to the minus (-)

OPERATING INSTRUCTIONS

CALIBRATE 2-WIRE TRANSMITTERS

+ I N - REF

Typical

 

2-Wire

(T/C / mA Shown)

Transmitter

CALIBRATE 4-WIRE TRANSMITTERS

To change the output value, press the UP or DOWN key for each digit. To ramp the output press and hold any UP or DOWN pushbutton. This function operates in all three output positions (HI, SET & LO). Press the HI & LO Quik-Cheks to recall stored outputs.

The TechChek 830-KP supplies a nominal 24 Volts DC at 24 mA to the 2-Wire transmitter. The current output of the transmitter will be accurately displayed by the 830-KP. Calibrate the Transmitter in the usual manner and disconnect the 830-KP.

+ I N - REF

Typical

4-Wire

Transmitter

Loop

Power Supply

24 VDC

30

31

Image 17
Contents Process Calibrator General Information ContentsConnections Rear Label Condensed Operation GuideGeneral TURN-ON Operating Instructions General Display Contrast PowerDisplay Backlighting Turn OFFStoring QUIK-CHEK Outputs Source ModeRecalling QUIK-CHEK Outputs Calibrate Milliamp InputsCalibrate Voltage Inputs Simulate 2-WIRE TransmittersWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Resistance Inputs Calibrate Thermocouple InputsThree Wire RTD Connection Two Wire RTD ConnectionFour Wire RTD Connection Calibrate RTD InputsCalibrate Frequency Inputs OUT of Range SignalsRead Functions MIN/MAXConnection for DC Volts to 200.0 vhi Read Milliamp OutputsRead DC and AC Voltage Outputs MV, V, Vhi and Vac Measure Thermocouple Sensors Read AC VoltagesCheck Continuity Read ResistanceOhms Measure RTD Sensors BlackCount Frequencies Connection for signals with amplitudes to 250 V RMSCalibrate 4-WIRE Transmitters Calibrate 2-WIRE TransmittersCalibrate Transmitters ANY Source Function Read mA, Read %, Read DP%Read Pressure Fitting InstallationRead Presssure Calibrate Pressure Transmitters Read Pressure Specifications Cold Junction Accuracy ±1C Measure DC Volts RangesRead Thermocouples Read OhmsType Range Accuracy Thermocouple SpecificationsRTD Type Alpha Range Accuracy RTD SpecificationsOrdering Information WarrantyPN 1800470 Rev a April