Altek 830-KP Measure DC Volts Ranges, Cold Junction Accuracy ±1C, Read Thermocouples, Read Ohms

Page 21

SPECIFICATIONS

MEASURE DC VOLTS

RANGES:

0.00to 110.00 mV; 0.00 to 10.25 V; 0.0 to 200.0 V

ACCURACY:

±(0.05% of 110 mV+ 0.01mV) = ±0.07 mV ±(0.05% of 10.25 V + 0.01V) = ±0.02V ±(2% of 200.0 V + 0.1V) = ±4.1 V

INPUT RESISTANCE: >1 Meg Ohm to 10.25V, >5 Meg Ohm to 200V

SOURCE RESISTANCE EFFECT: 0.01% per 100 Ohms

SOURCE THERMOCOUPLES

THERMOCOUPLE TYPES: J, K, T, E, N, R & S

RESOLUTION: 1°C or 1°F

ACCURACY: °C ±(0.05% OF 80 MV + 1°C); °F ±(0.05% OF 80 MV + 1°F)

COLD JUNCTION ACCURACY: ±1°C

COLD JUNCTION EFFECT: within 0.05°C per °C change

OUTPUT IMPEDANCE: <0.3 Ohms

SOURCE CURRENT: >10 mA, Max

READ THERMOCOUPLES

THERMOCOUPLE TYPES & ACCURACY: Same as for SOURCE T/C

RESOLUTION: 0.1°C or 0.1°F

COLD JUNCTION ACCURACY: ±1°C

COLD JUNCTION EFFECT: within 0.05°C per °C change

INPUT IMPEDANCE: > 1 Meg Ohm

OPEN THERMOCOUPLE DETECTION: 450 millisecond pulse. Nominal threshold, 10 K Ohms.

READ OHMS

RANGE OHMS: 0.0 to 1000.0 Ohms

ACCURACY: ±(0.05% of 1000.0 Ohms + 0.1 Ohm) = ±0.6 Ohms

EXCITATION CURRENT SUPPLIED: 1 mA, nominal

CONTINUITY CHECKING

TEST CURRENT: Nominal 1 mA

THRESHOLD: 100 Ohm ±20%

INDICATION: Steady tone & Symbol on LCD plus Ohm Reading

SPECIFICATIONS

SOURCE RTD & OHMS

RTD TYPES:

Pt 100for 1.3850 (DIN/IEC 751 & New JIS), 1.3902 (Burns), 1.3926 (US Lab) & 1.3916 (Old JIS 1604C-1981)

Ni 120, Cu 10& Cu 50RTD RESOLUTION: 1°C or 1°F RANGE OHMS: 0.0 to 400.0 Ohms

ACCURACY: ±0.05% of Full Scale + 0.075 mV/mA Excitation Current

ACCURACY OHMS: ±(0.05% of 400.0 Ohms + 0.1 Ohm = ±0.3 Ohms (At 1 mA Excitation Current)

TEMPERATURE EFFECT: ±((0.035 mV/°C)*(1/mAExcitation Current))

ALLOWABLE EXCITATION CURRENT: 0.125 to 2.0 mA continuous DC

READ RTD

RTD TYPES & RESOLUTION: Same as for SOURCE RTD

RTD RANGE (IN OHMS): 0.0 to 400.0 Ohms

RTD ACCURACY (OHMS): ±(0.05% of 400.0 Ohms + 0.1 Ohm = ±0.3 Ohms

EXCITATION CURRENT SUPPLIED: 1 mA, nominal

FREQUENCY SOURCE

RANGES: 1 to 1500 CPM (Count-Per-Minute); 1 to 1500 Hz, 0.01 to 15.00 kHz

ACCURACY:

±(0.05% of 1500 CPM + 1 CPM) = ±2 CPM ±(0.05% of 1500 Hz + 1 Hz) = ±2 Hz ±(0.05% of 15.00 kHz + 0.01 kHz) = ±0.02 kHz

OUTPUT WAVEFORM: Square Wave, Zero Crossing, -1V to +5V ±10%

RISETIME: Hz <5 microseconds; CPM <100 microseconds

OUTPUT IMPEDANCE: <100 Ohms

SOURCE CURRENT: >1 mA at 10 kHz

SHORT CIRCUIT DURATION: Infinite

MEASURE FREQUENCY

RANGES & ACCURACY: Same as for FREQUENCY SOURCE

TRIGGER LEVEL: 1 V RMS, DC coupled to 10.25 V; 7 V RMS, DC coupled to 250 V

INPUT IMPEDANCE: > 1Meg Ohm + 60 pF

38

39

Image 21
Contents Process Calibrator General Information ContentsRear Label Condensed Operation Guide General TURN-ONConnections Operating Instructions General Display Contrast PowerDisplay Backlighting Turn OFFStoring QUIK-CHEK Outputs Source ModeRecalling QUIK-CHEK Outputs Calibrate Milliamp InputsSimulate 2-WIRE Transmitters Wire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP %Calibrate Voltage Inputs Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Resistance Inputs Calibrate Thermocouple InputsThree Wire RTD Connection Two Wire RTD ConnectionFour Wire RTD Connection Calibrate RTD InputsCalibrate Frequency Inputs OUT of Range SignalsRead Functions MIN/MAXRead Milliamp Outputs Read DC and AC Voltage Outputs MV, V, Vhi and VacConnection for DC Volts to 200.0 vhi Measure Thermocouple Sensors Read AC VoltagesRead Resistance OhmsCheck Continuity Measure RTD Sensors BlackCount Frequencies Connection for signals with amplitudes to 250 V RMSCalibrate 4-WIRE Transmitters Calibrate 2-WIRE TransmittersCalibrate Transmitters ANY Source Function Read mA, Read %, Read DP%Fitting Installation Read PresssureRead Pressure Calibrate Pressure Transmitters Read Pressure Specifications Cold Junction Accuracy ±1C Measure DC Volts RangesRead Thermocouples Read OhmsType Range Accuracy Thermocouple SpecificationsRTD Type Alpha Range Accuracy RTD SpecificationsOrdering Information WarrantyPN 1800470 Rev a April