Altek 830-KP manual Operating Instructions

Page 4
RANGE LO SET HI
mA 4.00 12.00 20.00
mV 1.00 5.00 10.00
V 1.00 5.00 10.00 T/C J,T,E,K,N All points 0°C/32°F T/C R & S All points 538°C/1000°F
Ohms 100.0 200.0 400.0
RTD All points 0°C/32°F
kHz 1.00 5.00 10.00
Hz 100 500 1000
CPM 100 500 1000

OPERATING INSTRUCTIONS

GENERAL

FIELD & BENCH USE

TechChek 830-KP comes with a carrying case and a built-in tilt stand/hanger. The 830-KP is held securely in the case by VELCRO® for use with the carrying case open. The carrying case also has a snap- on belt loop which can be looped around a pipe or rail.

The tilt stand is easily raised by pulling the stand until it locks into place. The stand can also be reversed for use as a hanger to suspend the 830-KP.

CHANGING BATTERIES

Low battery is indicated by a battery symbol on the display. Approximately four hours of operation remain before the LCD blanks and TechChek 830-KP shuts itself down. Turn the 830-KP off, loosen the captive screw securing the battery compartment and lift off the cover from the bottom of the case. The six “AA” batteries are easily removed and replaced (alkaline supplied and recommended). Replace the battery compartment cover by inserting the tabs and tightening the screw.

OPERATING INSTRUCTIONS

GENERAL

CONFIGURING TEMPERATURE SCALES

The thermocouple and RTD ranges may be configured for full time use of °C, full time use of °F or selectable °C and °F operation. This configuration is part of the DEFAULT SETTINGS below.

AUTO-OFF

TechChek 830-KP can be set up to turn itself off after 30 minutes of inactivity. The internal timer is reset to 30 minutes each time a pushbutton is pressed.

DEFAULT SETTINGS

TechChek 830-KP may be restored to the factory default setting. This will reset the HI and LO “QUIK-CHEK” memories according to the table below and the SET memory to midrange between HI and LO.

1)Press and hold the STORE/RESET push-button while turning the 830-KP on.

2)Keep pressing the push-button until SETTING UP DEFAULT appears on the display then release the push-button.

3)°F/°C SETUP appears on the display.

4)Press an UP/DOWN key to select °F & °C, °F Only or °C Only. °C & °F is the default if no selection is made.

5)Press STORE/RESET push-button to store your choice or wait five seconds for the 830-KP to automatically store your choice and AUTO-OFF SETUP appears on the display.

6)Press an UP/DOWN key to select AUTO-OFF ENABLE. AUTO-OFF causes the 830-KP to turn itself off after 30 minutes of inactivity to preserve the batteries. Select AUTO-

OFF DISABLE for continuous QUIK-CHEK DEFAULTS operation.

7)Press STORE/RESET push-button to store your choice or wait five seconds for the 830-KP to automatically store your choice.

8)The firmware revision will be displayed, then the unit will begin

4

5

Image 4
Contents Process Calibrator Contents General InformationGeneral TURN-ON Rear Label Condensed Operation GuideConnections Operating Instructions Power General Display ContrastDisplay Backlighting Turn OFFSource Mode Storing QUIK-CHEK OutputsRecalling QUIK-CHEK Outputs Calibrate Milliamp InputsWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Simulate 2-WIRE TransmittersCalibrate Voltage Inputs Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Thermocouple Inputs Calibrate Resistance InputsTwo Wire RTD Connection Three Wire RTD ConnectionFour Wire RTD Connection Calibrate RTD InputsOUT of Range Signals Calibrate Frequency InputsRead Functions MIN/MAXRead DC and AC Voltage Outputs MV, V, Vhi and Vac Read Milliamp OutputsConnection for DC Volts to 200.0 vhi Read AC Voltages Measure Thermocouple SensorsOhms Read ResistanceCheck Continuity Black Measure RTD SensorsConnection for signals with amplitudes to 250 V RMS Count FrequenciesCalibrate 2-WIRE Transmitters Calibrate 4-WIRE TransmittersCalibrate Transmitters ANY Source Function Read mA, Read %, Read DP%Read Presssure Fitting InstallationRead Pressure Calibrate Pressure Transmitters Read Pressure Specifications Measure DC Volts Ranges Cold Junction Accuracy ±1CRead Thermocouples Read OhmsThermocouple Specifications Type Range AccuracyRTD Specifications RTD Type Alpha Range AccuracyWarranty Ordering InformationPN 1800470 Rev a April