OPERATING INSTRUCTIONS
READ MILLIAMP OUTPUTS
mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
Choose this function to measure from 0.00 to +24.00 milliamps.
1)Open the current loop at any convenient point along the signal path
2)Press the MILLIAMP/V
3)Press the DISPLAY/SOURCE/READ
4)Press the mA/%/% DP FLOW
5)Connect the red READ (+) lead of the calibrator to the more positive point of the break and the black READ lead
Display the present reading, Maximum or Minimum by pressing the READ, MAX or MIN
Milliamp Output Signal
Controller
Transmitter
P/I
DCS
OPERATING INSTRUCTIONS
READ DC and AC VOLTAGE OUTPUTS
mV, V, Vhi and Vac
Choose this function to measure from 0.00 to 110.00 millivolts (mV),
0.00to 10.25 DC Volts (V). Use the high voltage connection to read from 0.0 to 200.0 VDC (Vhi). See the next page for Volts AC.
1)Press the MILLIAMP/V
3)Press the DISPLAY/SOURCE/READ
4)Press the TYPE/ENG UNITS
5)Connect the red READ (+) lead and the black READ
Signals above or below those available for the currently selected range will be indicated by OVER and UNDER on the display.
Connection for millivolts(mV) and Volts below 10.25 VDC (V)
DC Voltage
Output Signal
Controller
Transmitter
Power Supply
Connection for DC Volts to 200.0 (vhi)
Voltage
Output Signal
Power Supply
Loop Voltage
20 | 21 |