OPERATING INSTRUCTIONS

READ AC VOLTAGES

Choose this function to measure from 0.0 to 250.0 V True RMS.

1)Press the MILLIAMP/V push-button until mV, V, Vhi or VAC appear on the display

3)Press the DISPLAY/SOURCE/READ push-button until READ appears on the display

4)Press the TYPE/ENG UNITS pushbuuton to select VAC on the display

5)Connect the red READ (+) lead and the black READ (-) lead of the calibrator across the voltage to be measured.

Signals above or below those available for the currently selected range will be indicated by OVER and UNDER on the display.

CAUTION: Care should be used when measuring AC voltage. The included safety test probes or safety alligator clips should be used. Do not exceed voltage limits shown on calibrator.

OPERATING INSTRUCTIONS

MEASURE THERMOCOUPLE SENSORS

Choose this function to read a thermocouple. The input of the TechChek 830-KP is automatically cold junction compensated.

1)Disconnect the thermocouple from any instrument.

2)Press the T/C/RTD/pushbutton until the word TYPE & any T/C appear on the display.

3)Press the DISPLAY/SOURCE/READ push-button until READ and any T/C TYPE appear on the display

4)Use the proper thermocouple wire and corresponding miniature T/C connector to connect TechChek 830-KP to the thermocouple.

To Change the Thermocouple type

1)Press the TYPE/ENG UNITS push-button. The words SELECT T/C TYPE and the current selection appear on the display.

2)Press any UP/DOWN push-button until the desired T/C type appears.

3)Press the STORE/RESET push-button or wait 5 seconds to store the selection

AC Voltage Signal

Mains

Line Voltage

+ -

22

23

Image 13
Contents Process Calibrator General Information ContentsGeneral TURN-ON Rear Label Condensed Operation GuideConnections Operating Instructions General Display Contrast PowerDisplay Backlighting Turn OFFStoring QUIK-CHEK Outputs Source ModeRecalling QUIK-CHEK Outputs Calibrate Milliamp InputsWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Simulate 2-WIRE TransmittersCalibrate Voltage Inputs Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Resistance Inputs Calibrate Thermocouple InputsThree Wire RTD Connection Two Wire RTD ConnectionFour Wire RTD Connection Calibrate RTD InputsCalibrate Frequency Inputs OUT of Range SignalsRead Functions MIN/MAXRead DC and AC Voltage Outputs MV, V, Vhi and Vac Read Milliamp OutputsConnection for DC Volts to 200.0 vhi Measure Thermocouple Sensors Read AC VoltagesOhms Read ResistanceCheck Continuity Measure RTD Sensors BlackCount Frequencies Connection for signals with amplitudes to 250 V RMSCalibrate 4-WIRE Transmitters Calibrate 2-WIRE TransmittersCalibrate Transmitters ANY Source Function Read mA, Read %, Read DP%Read Presssure Fitting InstallationRead Pressure Calibrate Pressure Transmitters Read Pressure Specifications Cold Junction Accuracy ±1C Measure DC Volts RangesRead Thermocouples Read OhmsType Range Accuracy Thermocouple SpecificationsRTD Type Alpha Range Accuracy RTD SpecificationsOrdering Information WarrantyPN 1800470 Rev a April