Altek 830-KP manual Thermocouple Specifications, Type Range Accuracy

Page 22

THERMOCOUPLE SPECIFICATIONS

T/C

°C

 

°F

 

TYPE

RANGE ACCURACY

RANGE

ACCURACY

 

 

 

 

 

J

100 to 1200

±1.7

212 to 2192

±2.3

 

-50 to 99

±1.8

-58 to 211

±2.5

 

-150 to -49

±2.2

-238 to -57

±3.2

 

-200 to -149

±2.8

-328 to -237

±4.3

K

1100 to 1372

±2.2

2012 to 2500

±3.1

 

0 to 1099

±2.0

32 to 2011

±2.9

 

-100 to -1

±2.3

-148 to 31

±3.3

 

-200 to -99

±3.6

-328 to -147

±5.7

 

 

 

 

 

T

200 to 400

±1.7

392 to 752

±2.4

 

0 to 199

±1.9

32 to 391

±2.7

 

-100 to -1

±2.4

-148 to 31

±3.5

 

-200 to -99

±3.5

-328 to -147

±5.6

 

 

 

 

 

E

250 to 1000

±1.5

482 to 1832

±2.0

 

50 to 249

±1.6

122 to 481

±2.1

 

-100 to 49

±1.9

-148 to 121

±2.6

 

-200 to -99

±2.6

-328 to -147

±3.9

 

 

 

 

 

THERMOCOUPLE SPECIFICATIONS

T/C

°C

 

°F

 

TYPE

RANGE ACCURACY

RANGE

ACCURACY

 

 

 

 

 

N

300 to 1300

±2.1

572 to 2372

±3.0

 

100 to 299

±2.3

212 to 571

±3.3

 

-50 to 99

±2.6

-58 to 211

±4.0

 

-200 to -49

±5.0

-328 to -57

±8.2

R

1750 to 1768

±4.2

3182 to 3214

±6.9

950 to 1749

±4.0

1742 to 3181

±6.5

 

 

650 to 949

±4.4

1202 to 1741

±7.2

 

300 to 649

±5.1

572 to 1201

±8.4

 

 

 

 

 

S

1700 to 1768

±4.9

3092 to 3214

±8.0

1050 to 1699

±4.4

1922 to 3091

±7.1

 

 

700 to 1049

±4.8

1292 to 1921

±7.8

 

300 to 699

±5.4

572 to 1291

±8.9

 

 

 

 

 

Note: Thermocouple accuracies are based on an 80.00 mV Span T/C Accuracy for °C is ±(0.05% of 80.00 mV + 1°C)

T/C Accuracy for °F is ±(0.05% of 80.00 mV + 1°F)

Source resolution is 1 °C or °F. Read resolution is 0.1 °C or °F

40

41

Image 22
Contents Process Calibrator Contents General InformationGeneral TURN-ON Rear Label Condensed Operation GuideConnections Operating Instructions Display Backlighting PowerGeneral Display Contrast Turn OFFRecalling QUIK-CHEK Outputs Source ModeStoring QUIK-CHEK Outputs Calibrate Milliamp InputsWire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP % Simulate 2-WIRE TransmittersCalibrate Voltage Inputs Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Thermocouple Inputs Calibrate Resistance InputsFour Wire RTD Connection Two Wire RTD ConnectionThree Wire RTD Connection Calibrate RTD InputsRead Functions OUT of Range SignalsCalibrate Frequency Inputs MIN/MAXRead DC and AC Voltage Outputs MV, V, Vhi and Vac Read Milliamp OutputsConnection for DC Volts to 200.0 vhi Read AC Voltages Measure Thermocouple SensorsOhms Read ResistanceCheck Continuity Black Measure RTD SensorsConnection for signals with amplitudes to 250 V RMS Count FrequenciesCalibrate Transmitters ANY Source Function Calibrate 2-WIRE TransmittersCalibrate 4-WIRE Transmitters Read mA, Read %, Read DP%Read Presssure Fitting InstallationRead Pressure Calibrate Pressure Transmitters Read Pressure Specifications Read Thermocouples Measure DC Volts RangesCold Junction Accuracy ±1C Read OhmsThermocouple Specifications Type Range AccuracyRTD Specifications RTD Type Alpha Range AccuracyWarranty Ordering InformationPN 1800470 Rev a April