OPERATING INSTRUCTIONS

CALIBRATE THERMOCOUPLE INPUTS

Choose this function to simulate a thermocouple signal into any instrument requiring a thermocouple input. The output of the 830- KP is automatically cold junction compensated.

1)Disconnect the thermocouple from the instrument being calibrated.

2)Press the T/C/RTD/pushbutton until the word TYPE & any T/C appear on the display.

3)Press the DISPLAY/SOURCE/READ push-button until SOURCE and any T/C TYPE appear on the display

4)Use the proper thermocouple wire and corresponding miniature T/C connector to connect TechChek 830-KP to the thermocouple.

Output temperature is continuously adjustable with the UP/DOWN push-buttons. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs”.

To Change the Thermocouple type

1)Press the TYPE/ENG UNITS push-button. The words SELECT T/C TYPE and the current selection appear on the display.

2)Press any UP/DOWN push-button until the desired T/C type appears.

3)Press the STORE/RESET push-button or wait 5 seconds to store the selection

+-

Instrument with

Thermocouple Input

Controller

Temperature Indicator

Temperature Trip or Alarm

Temperature Transmitter

14

OPERATING INSTRUCTIONS

CALIBRATE RESISTANCE INPUTS

Choose this function to simulate a resistance into a variety of instruments.

1)Disconnect one or both input wires from the device to be calibrated

2)Press the T/C/RTD/pushbutton until the symbol appears on the display.

3)Press the DISPLAY/SOURCE/READ push-button until SOURCE appears on the display

4)Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus (-).

Output resistance is continuously adjustable with the UP/DOWN push-buttons. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs”.

Receiver Input

Ohms or 2-Wire RTD

Controller

Transmitter

Computer

15

Image 9
Contents Process Calibrator General Information ContentsRear Label Condensed Operation Guide General TURN-ONConnections Operating Instructions General Display Contrast PowerDisplay Backlighting Turn OFFStoring QUIK-CHEK Outputs Source ModeRecalling QUIK-CHEK Outputs Calibrate Milliamp InputsSimulate 2-WIRE Transmitters Wire mA, 2-WIRE % Percent of 4 to 20 mA 2-WIRE DP %Calibrate Voltage Inputs Check 1-5 Volt Inputs Without Disconnecting Wires Calibrate Resistance Inputs Calibrate Thermocouple InputsThree Wire RTD Connection Two Wire RTD ConnectionFour Wire RTD Connection Calibrate RTD InputsCalibrate Frequency Inputs OUT of Range SignalsRead Functions MIN/MAXRead Milliamp Outputs Read DC and AC Voltage Outputs MV, V, Vhi and VacConnection for DC Volts to 200.0 vhi Measure Thermocouple Sensors Read AC VoltagesRead Resistance OhmsCheck Continuity Measure RTD Sensors BlackCount Frequencies Connection for signals with amplitudes to 250 V RMSCalibrate 4-WIRE Transmitters Calibrate 2-WIRE TransmittersCalibrate Transmitters ANY Source Function Read mA, Read %, Read DP%Fitting Installation Read PresssureRead Pressure Calibrate Pressure Transmitters Read Pressure Specifications Cold Junction Accuracy ±1C Measure DC Volts RangesRead Thermocouples Read OhmsType Range Accuracy Thermocouple SpecificationsRTD Type Alpha Range Accuracy RTD SpecificationsOrdering Information WarrantyPN 1800470 Rev a April