The input port(s) on the instrument are designed to accept a hose with 4mm I.D. and 1mm wall. The hose is fitted by unscrewing the clamping nut, sliding the hose on to the stepped shaft and refitting the nut.

Note: Do not tighten fittings attached to the pressure ports to a torque greater than 2Nm otherwise the pressure seals may be damaged.

HAND PUMP OPERATION (if applicable)

Use the T piece to connect the hand pump to the instrument’s + port and to the equipment being tested. Exhaust the system by turning the release valve on the hand pump anti-clockwise and set the fine adjustment threaded piston to mid range.

Close the release valve and apply pump action to the hand pump piston to generate the required approximate pressure. Fine adjustment of the desired pressure can then be made by rotating the threaded piston. The release valve may be used to vent the system. Note: The hand pump is capable of generating 7bar max. It is, therefore, essential that extra care is taken not to over-pressure the instrument.

A pump must not be used on models HHP-2000, 2020 and 2080

PLUS FUNCTIONS

Models HHP-2020, 2021, 2022, 2023, 2024, 2025, 2080, 2081, 2082, 2084, 2085

SMOOTHING MODE

Press = key and the instrument continuously averages the last four measurements to give a more stable reading on the display. Press = key again to return to normal measurement.

UNITS OF PRESSURE

Press key to select required units of pressure. Chosen units will be saved when the instrument is switched off.

Please refer to the table on the back page.

The HHP-20x0 0-25mbar instruments cannot show the µbar or mmHg symbols on their displays. When switching the units on they are reading in µbar and so show no unit symbol. Pushing the autorange button will put the unit into mbar and show the mbar symbol. When reading in mmHg only the symbol Hg will show on the display.

This function will not operate while in store mode. Note: HHP-2024 - 2084

K = x 103 when using in H2O and Pa. M = x 106 when using in H2O and Pa.

MAX/MIN RECORDER

Press

key to start. The instrument will display alternating

 

symbols and current reading. Press again to display the

maximum pressure reading and

symbol will be displayed.

Press again to display the minimum pressure reading and

symbol will be displayed. Press again to return to the actual pressure display.

Press X key to display the average pressure since the MAX/MIN

mode was activated and symbol will appear on display.

Press X key again to return to actual pressure display.

Image 4
Contents HHP-2000 Series Manometer Operating Instructions Features BatteriesIP65/IP67 Rating On OFFBacklight DisplayOUT of Range ZeroingSmoothing Mode MAX/MIN RecorderHose Fitting Plus FunctionsSetting Internal Clock Preset Interval LoggingRange Lock Advanced FunctionsLeak Testing Setting UPOutput of Data AveragingWiring instructions Pressure Transducer SelectionModels HHP-2024 User Defined ScaleBedienung Batterie EIN/AUSSCHALTENModelle Meßbereich Bereichsüberschreitung Besondere MerkmaleHintergrundbeleuchtung AnzeigeBereichsüberschreitung NullpunktkorrekturMAX/MIN Messung Automatische Bereichseinstellung Zusätzliche FunktionenAUFZEICHNUNGS-MODUS Zeit EinstellenDatenausgabe MITTELWERT-MESSUNGENLECKSUCH-MESSUNGEN EinstellungenGenauigeiten DRUCKWANDLER-AUSWAHLExterner DRUCKWANDLER, Anschluss Definierbare SkalaPilas Protección IP65/IP67Encendido Apagado GamaRetroiluminacón Registro DE MÁXIMO/MÍNIMO Instalación DE TuboFunciones Adicionales Modo DE AlisamientoBloqueo DE Rango Funciones AvanzadasRegistro Logging Ajuste DEL Reloj InternoRegistro a Intervalos Predeterminados Salida DE DatosPromedio Verificación DE FugasConfiguración Realización DE UN EnsayoSelección DEL Transductor DE Presión Modelos HHP-2024Escala Definida POR EL Usuario Exactitud DE LOS InstrumentosBatterie Niveau IP65/IP67ALLUMER/ARRÊTER L’APPAREIL Gamme DE MesuresAffichage Valeur MINIMALE/MAXIMALE Bloquer LA Résolution Fonctions Plus AvancéesEnregistrer Comment Régler L’HORLOGE InterneMode Moyen Enregistrement PrérégléSortie DE Données Détection DE FuiteMise EN Marche Faire UN TestSélection DU Capteur DE Pression Modèles HHP-2024Précision Echelle PersonnaliseeCalibration Valutazione IP65/IP67 PortataModelli Portata di Funzionamento Portata in eccesso CaratteristicheRetroilluminazione Fuori PortataAzzeramento Azzeramento AssolutoRegistratore MAX/MIN Sistemazione Tubo FlessibileFunzioni Plus Modalità DI StabilizzazioneEstensione Dispositivo DI Bloccaggio Funzioni D’AVANGUARDIARegistrazione Regolazione DELL’OROLOGIO InternoUscita Dati Registrazione SU RichiestaFunzione Intermedia Prova Tenuta Acqua PredeterminazioniProva DI Prestazione Selezione DEL Trasduttore DI PressioneScala Definita DALL’UTENTE Istruzioni di cablaggioPrecisione Dello Strumento Repair instructions and/or specific problems relative to Warranty / DisclaimerReturn Requests / Inquiries Servicing North America For immediate technical or application assistanceServicing Europe USAServicing USA and Canada Call Omega Toll Free USACall for your Free Handbook Request Form today 203 359-RUSH Page Page Reading Input PSI