Omega CNiS8C, CNiS8DH Strain Gauge, Figure below shows the wiring hookup for 4-wire bridge input

Page 13

2.3.4 Strain Gauge

The figure below shows the wiring hookup for 4-wire bridge input.

Figure 2.7

a)4-Wire Voltage/Bridge Input with Internal Excitation Wiring Hookup

b)4-Wire Bridge Input with External Excitation Wiring Hookup

In 4-Wire connections the voltage drop across long excitation lead wires of strain gauge bridge may cause measurement errors. The output of a strain gauge bridge also depends on the stability of excitation voltage. To correct for voltage drop and changes in excitation voltage, 6-wire input configuration and ratio measurement are used.

In order for the Ratiometric to work properly, the External Excitation should not drop below 4.6 Vdc.

The figure below shows 6-wire hookup for 6-wire bridge input.

Figure 2.8

a)6-Wire Bridge Input with Internal Excitation and Ratio Measurement Wiring Hookup

b)6-Wire Bridge Input with External Excitation and Ratio Measurement Wiring Hookup

9

Image 13
Contents CNiS8, CNiS8C CNiS8DH, CNiS8DV CNiS16, CNiS16D, CNiS32 Servicing North America Table of Contents List of Tables List of FiguresTIP Description Part IntroductionEMC Considerations Safety ConsiderationsBefore You Begin Front Panel Part SetupFront Panel Annunciators Rear Panel ConnectionsInput PowerOutput OptionMain Power Connections Fuse Requirements Electrical Installation Power ConnectionsProcess Current Strain Gauge Figure below shows the wiring hookup for 4-wire bridge inputOutputs Wiring Hookup 11 Snubber Circuits Wiring HookupRS-232 Output Wiring Hookup b RS-485 Output Wiring Hookup Wiring OutputsTop View Location of S2 on 1/8 DIN Compact Unit Excitation OutputJumper Connections Turning your Controller On for the First Time Part Operation Configuration ModeFlow Chart for ID and Set Points Menu Menu ConfigurationEntering Your NON-DEFAULT SETPOINT/ID Security ID Number Entering Your NON-DEFAULT Full Security ID NumberID Number Menu Setpoint Set Points MenuEnter Configuration Menu Configuration MenuInput Type Menu Enter Input Type Menu INPUT/READING Resolution SubmenuInput Type Menu Ratiometric Operation SubmenuButton Selection Submenu Reading Configuration MenuDecimal Point Submenu Enter Reading Configuration MenuKNOWN/UNKNOWN Loads Submenu Linearization Points SubmenuFilter Constant Submenu Example 1 Scaling with Known Loads On-Line Calibration Input/Reading Scale and Offset MenuConversion Table Input Resolution MultiplierExample 2 Scaling without Known Loads Page Page Enter Alarm 1 Menu Alarm 1 MenuAlarm 1 ABSOLUTE/DEVIATION Submenu Alarm 1 ENABLE/DISABLE SubmenuAlarm 1 LATCH/UNLATCH Submenu Active Submenu Contact Closure SubmenuAlarm 1 LOW Value Submenu Alarm ENABLE/DISABLE AT Power onAlarm 1 HI Value Submenu Enter Analog Output Menu Analog Output Retransmission MenuCURRENT/VOLTAGE Submenu Analog Output ENABLE/DISABLE SubmenuReading OUTFor example Enter Alarm 2 Menu Alarm 2 MenuLoop Break Time Menu Alarm 2 ENABLE/DISABLE SubmenuEnter Loop Break Time Menu Setpoint Deviation ENABLE/DISABLE Submenu Loop Break ENABLE/DISABLE SubmenuLoop Break Time Value Submenu Output 1 Menu Flow Chart for Output 1 MenuSelf Submenu Enter Output 1 MenuMINIMUM/PERCENT LOW Submenu Control Type Output MAXIMUM/PERCENT High SubmenuAuto PID Submenu Action Type SubmenuAnti Integral Submenu Proportional Band Submenu Start Auto Tune PIDRate Setup Submenu Reset Setup SubmenuCycle Time Submenu Deadband Submenu Damping Factor SubmenuEnter Output 2 Menu Output 2 MenuControl Type Submenu Be called Reverse Acting Cycle Time Submenu Ramp & Soak Menu Ramp ENABLE/DISABLE SubmenuEnter Ramp and Soak Menu Ramp Value Submenu Soak ENABLE/DISABLE SubmenuSoak Value Submenu ID Code Menu Entering or Changing Your NON-DEFAULT ID CodeEnter ID Code Menu Full Security Level Submenu Entering or Changing Your Default ID CodeSETPOINT/ID Security Level Submenu Communication Option Menu 13 Flow Chart for Communication Option MenuCommunication Parameters Submenu Enter Communication Option MenuBaud Submenu Parity SubmenuBUS Format Submenu Stop BIT SubmenuModbus Protocol Submenu Line Feed SubmenuCommunication Interface Standard Submenu Data Flow Mode SubmenuData Separation Character Submenu Data Format Submenu Alarm Status SubmenuMain Reading Submenu Peak Value SubmenuUnit Submenu not applicable Address Setup SubmenuAddress Value Submenu Transmit Time Interval SubmenuDisplay Color Selection Menu Alarm 1 Display Color SubmenuEnter Display Color Selection Menu Normal Color Display SubmenuExample Alarm 2 Display Color SubmenuDisplay Colors change sequences Green RED AmberAmber Green Control Part SpecificationsSSR Control Output 1Excitation InsulationGeneral Factory preset value Part Factory Preset ValuesMenu Items Factory Preset Values PID Measurement Category Part CE Approvals InformationPage Page Return REQUESTS/INQUIRIES WARRANTY/DISCLAIMERPRESSURE, Strain and Force TemperatureFLOW/LEVEL Data Acquisition