Omega CNiS8DH, CNiS32, CNiS16D, CNiS8DV Enter Input Type Menu, Ratiometric Operation Submenu

Page 22

ENTER INPUT TYPE MENU:

Press a 1) Press a, if necessary, until CNFG prompt appears.

Press d 2) Display advances to INPT Input Menu.

Press d 3) Display flashes 0-0.1, 0-1.0, 0-10 or 0-20 (0 to 100 mV, 0 to 1 V, 0 to 10 V or 0 to 20 mA).

INPUT TYPE MENU:

Press b 4) Scroll through the available selection of input ranges 0-0.1, 0-1.0, 0-10 or 0-20 to the selection of your choice.

Press d 5) Display shows STRD stored message momentarily and then advances to the RT1B Ratiometric Operation Submenu.

Input Types:

100 mV

1 V

10 V

0 – 20 mA

Display:

0-0.1

0-1.0

0-10

0-20

RATIOMETRIC OPERATION SUBMENU:

Press d 6) Display flashes previous selection of ENBL Enable or DSBL Disable.

Press b 7) Scroll through the available selection ENBL or DSBL (flashing).

Press d 8) Display shows STRD stored message momentarily and then advances to RESO Input/Reading Resolution Submenu.

The Ratiometric operations are typically used for Strain gauge controller. If your controller is configured as Process (Voltage and Current), set RTIB to DSBL disable Ratiometric operations.

If ENBL Ratiometric operations Enabled was selected, the changes to the excitation voltage will be compensated through Ratio measurement. If DSBL Ratiometric operation Disabled was selected, any changes to the excitation voltage will effect the output of strain gauge bridge and, as a result, a reading of the controller.

INPUT/READING RESOLUTION SUBMENU:

Press a 9) Display flashes previous selection of LO Low or HI High resolution.

Press d 10) Scroll through the available selection LO or HI (flashing).

Press d 11) Display shows STRD stored message momentarily and then advances to BUTN Button Selection Submenu.

If LO Low Resolution was selected the resolution of the display is 10 µV. If HI High Resolution was selected the resolution of the display is 1 µV. In case of High Resolution, the maximum input signal is 10 mV.

18

Image 22
Contents CNiS8, CNiS8C CNiS8DH, CNiS8DV CNiS16, CNiS16D, CNiS32 Servicing North America Table of Contents List of Figures List of TablesTIP Part Introduction DescriptionSafety Considerations EMC ConsiderationsBefore You Begin Front Panel Annunciators Part SetupFront Panel Rear Panel ConnectionsOutput PowerInput OptionElectrical Installation Power Connections Main Power Connections Fuse RequirementsProcess Current Figure below shows the wiring hookup for 4-wire bridge input Strain Gauge11 Snubber Circuits Wiring Hookup Outputs Wiring HookupWiring Outputs RS-232 Output Wiring Hookup b RS-485 Output Wiring HookupTop View Location of S2 on 1/8 DIN Compact Unit Excitation OutputJumper Connections Part Operation Configuration Mode Turning your Controller On for the First TimeMenu Configuration Flow Chart for ID and Set Points MenuEntering Your NON-DEFAULT SETPOINT/ID Security ID Number Entering Your NON-DEFAULT Full Security ID NumberID Number Menu Set Points Menu SetpointEnter Configuration Menu Configuration MenuInput Type Menu Input Type Menu INPUT/READING Resolution SubmenuEnter Input Type Menu Ratiometric Operation SubmenuReading Configuration Menu Button Selection SubmenuKNOWN/UNKNOWN Loads Submenu Enter Reading Configuration MenuDecimal Point Submenu Linearization Points SubmenuFilter Constant Submenu Input/Reading Scale and Offset Menu Example 1 Scaling with Known Loads On-Line CalibrationConversion Table Input Resolution MultiplierExample 2 Scaling without Known Loads Page Page Alarm 1 Menu Enter Alarm 1 MenuAlarm 1 ABSOLUTE/DEVIATION Submenu Alarm 1 ENABLE/DISABLE SubmenuAlarm 1 LATCH/UNLATCH Submenu Contact Closure Submenu Active SubmenuAlarm 1 LOW Value Submenu Alarm ENABLE/DISABLE AT Power onAlarm 1 HI Value Submenu Analog Output Retransmission Menu Enter Analog Output MenuReading Analog Output ENABLE/DISABLE SubmenuCURRENT/VOLTAGE Submenu OUTFor example Alarm 2 Menu Enter Alarm 2 MenuLoop Break Time Menu Alarm 2 ENABLE/DISABLE SubmenuEnter Loop Break Time Menu Setpoint Deviation ENABLE/DISABLE Submenu Loop Break ENABLE/DISABLE SubmenuLoop Break Time Value Submenu Flow Chart for Output 1 Menu Output 1 MenuSelf Submenu Enter Output 1 MenuMINIMUM/PERCENT LOW Submenu MAXIMUM/PERCENT High Submenu Control Type OutputAuto PID Submenu Action Type SubmenuAnti Integral Submenu Start Auto Tune PID Proportional Band SubmenuRate Setup Submenu Reset Setup SubmenuCycle Time Submenu Damping Factor Submenu Deadband SubmenuEnter Output 2 Menu Output 2 MenuControl Type Submenu Be called Reverse Acting Cycle Time Submenu Ramp & Soak Menu Ramp ENABLE/DISABLE SubmenuEnter Ramp and Soak Menu Ramp Value Submenu Soak ENABLE/DISABLE SubmenuSoak Value Submenu ID Code Menu Entering or Changing Your NON-DEFAULT ID CodeEnter ID Code Menu Full Security Level Submenu Entering or Changing Your Default ID CodeSETPOINT/ID Security Level Submenu 13 Flow Chart for Communication Option Menu Communication Option MenuBaud Submenu Enter Communication Option MenuCommunication Parameters Submenu Parity SubmenuModbus Protocol Submenu Stop BIT SubmenuBUS Format Submenu Line Feed SubmenuCommunication Interface Standard Submenu Data Flow Mode SubmenuData Separation Character Submenu Main Reading Submenu Alarm Status SubmenuData Format Submenu Peak Value SubmenuAddress Value Submenu Address Setup SubmenuUnit Submenu not applicable Transmit Time Interval SubmenuEnter Display Color Selection Menu Alarm 1 Display Color SubmenuDisplay Color Selection Menu Normal Color Display SubmenuDisplay Colors change sequences Alarm 2 Display Color SubmenuExample Green RED AmberAmber Green Part Specifications ControlExcitation Control Output 1SSR InsulationGeneral Part Factory Preset Values Factory preset valueMenu Items Factory Preset Values PID Part CE Approvals Information Measurement CategoryPage Page WARRANTY/DISCLAIMER Return REQUESTS/INQUIRIESFLOW/LEVEL TemperaturePRESSURE, Strain and Force Data Acquisition