Omega CNiS8C Conversion Table, Input Resolution Multiplier, Example 2 Scaling without Known Loads

Page 27

Max scale should not be more than 50% FS because of noise related issues.

Press d 27) Display flashes STRD stored message momentarily and then advances to ALR1 only, if a change was made, otherwise advances to ALR1 Alarm 1 Menu.

Example 2: Scaling without Known Loads.

If DSBL Disabled Load Submenu was selected, instrument is ready for scaling Without Known Loads method.

To scale without known inputs, calculate inputs based on transducer specifications and manually enter them on the via front panel push-buttons. The following example assumes load cells with this specification:

Maximum Load:

100.0 lb

Output:

3.0 mV/V

Sensor Excitation

10 V

Maximum Sensor Output = 3.0 (mV/V) x 10 (V) = 30 mV

1.Determine the correct values for Inputs (IN!1 and IN!2). Calculate IN!1 and IN!2 using the following equation:

IN = (Sensor Output) x (Converison Number) x (Multiplier)

Conversion number is a coefficient of conversion between input values and real full display range (10000 counts). See Table 3.1 below for proper conversion number.

Table 3.2 Conversion Table

INPUT RANGE

CONVERSION NUMBER

0

~ 100 mV

10000

/ (100 x 1) = 100 cts/mV

0

~ 1 V

10000

/ (1000 x 1) = 10 cts/mV

0

~ 10 V

10000

/ (1000 x 10) = 1 cts/mV

 

 

 

 

0

~ 20 mA

10000

/ (20 x 1) = 500 cts/mV

Multiplier determined by the Input Resolution setting (RESO in the INPT Menu). See Table 3.2 below for proper multiplier.

Table 3.3 Input Resolution Multiplier

INPUT RANGE

RESOLUTION

 

 

LOW

HIGH

0

~ 100 mV

1.0

10.0

0

~ 1 V

1.0

10.0

0

~ 10 V

1.0

10.0

0

~ 20 mA

1.0

10.0

23

Image 27
Contents CNiS8, CNiS8C CNiS8DH, CNiS8DV CNiS16, CNiS16D, CNiS32 Servicing North America Table of Contents List of Tables List of FiguresTIP Description Part IntroductionEMC Considerations Safety ConsiderationsBefore You Begin Rear Panel Connections Part SetupFront Panel Front Panel AnnunciatorsOption PowerInput OutputMain Power Connections Fuse Requirements Electrical Installation Power ConnectionsProcess Current Strain Gauge Figure below shows the wiring hookup for 4-wire bridge inputOutputs Wiring Hookup 11 Snubber Circuits Wiring HookupRS-232 Output Wiring Hookup b RS-485 Output Wiring Hookup Wiring OutputsExcitation Output Top View Location of S2 on 1/8 DIN Compact UnitJumper Connections Turning your Controller On for the First Time Part Operation Configuration ModeFlow Chart for ID and Set Points Menu Menu ConfigurationEntering Your NON-DEFAULT Full Security ID Number Entering Your NON-DEFAULT SETPOINT/ID Security ID NumberID Number Menu Setpoint Set Points MenuConfiguration Menu Enter Configuration MenuInput Type Menu Ratiometric Operation Submenu INPUT/READING Resolution SubmenuEnter Input Type Menu Input Type MenuButton Selection Submenu Reading Configuration MenuLinearization Points Submenu Enter Reading Configuration MenuDecimal Point Submenu KNOWN/UNKNOWN Loads SubmenuFilter Constant Submenu Example 1 Scaling with Known Loads On-Line Calibration Input/Reading Scale and Offset MenuInput Resolution Multiplier Conversion TableExample 2 Scaling without Known Loads Page Page Enter Alarm 1 Menu Alarm 1 MenuAlarm 1 ENABLE/DISABLE Submenu Alarm 1 ABSOLUTE/DEVIATION SubmenuAlarm 1 LATCH/UNLATCH Submenu Active Submenu Contact Closure SubmenuAlarm ENABLE/DISABLE AT Power on Alarm 1 LOW Value SubmenuAlarm 1 HI Value Submenu Enter Analog Output Menu Analog Output Retransmission MenuOUT Analog Output ENABLE/DISABLE SubmenuCURRENT/VOLTAGE Submenu ReadingFor example Enter Alarm 2 Menu Alarm 2 MenuAlarm 2 ENABLE/DISABLE Submenu Loop Break Time MenuEnter Loop Break Time Menu Loop Break ENABLE/DISABLE Submenu Setpoint Deviation ENABLE/DISABLE SubmenuLoop Break Time Value Submenu Output 1 Menu Flow Chart for Output 1 MenuEnter Output 1 Menu Self SubmenuMINIMUM/PERCENT LOW Submenu Control Type Output MAXIMUM/PERCENT High SubmenuAction Type Submenu Auto PID SubmenuAnti Integral Submenu Proportional Band Submenu Start Auto Tune PIDReset Setup Submenu Rate Setup SubmenuCycle Time Submenu Deadband Submenu Damping Factor SubmenuOutput 2 Menu Enter Output 2 MenuControl Type Submenu Be called Reverse Acting Cycle Time Submenu Ramp ENABLE/DISABLE Submenu Ramp & Soak MenuEnter Ramp and Soak Menu Soak ENABLE/DISABLE Submenu Ramp Value SubmenuSoak Value Submenu Entering or Changing Your NON-DEFAULT ID Code ID Code MenuEnter ID Code Menu Entering or Changing Your Default ID Code Full Security Level SubmenuSETPOINT/ID Security Level Submenu Communication Option Menu 13 Flow Chart for Communication Option MenuParity Submenu Enter Communication Option MenuCommunication Parameters Submenu Baud SubmenuLine Feed Submenu Stop BIT SubmenuBUS Format Submenu Modbus Protocol SubmenuData Flow Mode Submenu Communication Interface Standard SubmenuData Separation Character Submenu Peak Value Submenu Alarm Status SubmenuData Format Submenu Main Reading SubmenuTransmit Time Interval Submenu Address Setup SubmenuUnit Submenu not applicable Address Value SubmenuNormal Color Display Submenu Alarm 1 Display Color SubmenuDisplay Color Selection Menu Enter Display Color Selection MenuGreen RED Amber Alarm 2 Display Color SubmenuExample Display Colors change sequencesAmber Green Control Part SpecificationsInsulation Control Output 1SSR ExcitationGeneral Factory preset value Part Factory Preset ValuesMenu Items Factory Preset Values PID Measurement Category Part CE Approvals InformationPage Page Return REQUESTS/INQUIRIES WARRANTY/DISCLAIMERData Acquisition TemperaturePRESSURE, Strain and Force FLOW/LEVEL