Omega CNiS8DH, CNiS16, CNiS32 Communication Interface Standard Submenu, Data Flow Mode Submenu

Page 57

Press d

Press b

Press d

25)Display flashes previous selection for “Echo”.

26)Scroll through the available selections: NO, YES.

27)Display flashes STRD stored message momentarily and then advances to STND only if it was changed, otherwise press a to advance to STND Communication Standard Submenu.

COMMUNICATION INTERFACE STANDARD SUBMENU:

Determines whether device should be connected to an RS-232C serial port (as is commonly used on IBM PC-compatible computers) or via an RS-485 bus connected through appropriate RS-232/485 converter. When used in RS-485 Mode, the device must be accessed with an appropriate Address Value as selected in the Address Submenu described later.

Press d

Press b

Press d

28)Display flashes previous selection for “Standard”.

29)Scroll through the available selections: 232C, 485.

30)Display shows STRD stored message momentarily and then advances to MoDE only, if it was changed, otherwise press a to advance to MoDE Data Flow Mode Submenu.

DATA FLOW MODE SUBMENU:

Determines whether the instrument will wait for commands and data requests from the Serial Bus or whether the instrument will send data automatically and continuously to the Serial Bus. Devices configured for the RS-485 Communications Standard operate properly only under Command Mode.

Press d Press b

Press d

31)Display flashes previous selection for “Mode”.

32)Scroll through the available selections: CMD_ “Command”, CoNT “Continuous”.

33)Display shows STRD stored message momentarily and then advances to SEPR only, if it was changed, otherwise press a to advance to SEPR Data Separation Submenu.

DATA SEPARATION CHARACTER SUBMENU:

Determines whether data sent from the device in Continuous Data Flow Mode will be separated by spaces or by Carriage Returns.

Press d Press b

Press d

34)Display flashes previous selection for “Separation” Submenu.

35)Scroll through the available selections: SPCE “Space” or

_çR_ “Carriage Return”.

36)Display shows STRD stored message momentarily and then advances to DAT.F only, if it was changed, otherwise press a to advance to DAT.F Data Format Submenu.

53

Image 57
Contents CNiS8, CNiS8C CNiS8DH, CNiS8DV CNiS16, CNiS16D, CNiS32 Servicing North America Table of Contents List of Tables List of FiguresTIP Description Part IntroductionEMC Considerations Safety ConsiderationsBefore You Begin Front Panel Part SetupFront Panel Annunciators Rear Panel ConnectionsInput PowerOutput OptionMain Power Connections Fuse Requirements Electrical Installation Power ConnectionsProcess Current Strain Gauge Figure below shows the wiring hookup for 4-wire bridge inputOutputs Wiring Hookup 11 Snubber Circuits Wiring HookupRS-232 Output Wiring Hookup b RS-485 Output Wiring Hookup Wiring OutputsExcitation Output Top View Location of S2 on 1/8 DIN Compact UnitJumper Connections Turning your Controller On for the First Time Part Operation Configuration ModeFlow Chart for ID and Set Points Menu Menu ConfigurationEntering Your NON-DEFAULT Full Security ID Number Entering Your NON-DEFAULT SETPOINT/ID Security ID NumberID Number Menu Setpoint Set Points MenuConfiguration Menu Enter Configuration MenuInput Type Menu Enter Input Type Menu INPUT/READING Resolution SubmenuInput Type Menu Ratiometric Operation SubmenuButton Selection Submenu Reading Configuration MenuDecimal Point Submenu Enter Reading Configuration MenuKNOWN/UNKNOWN Loads Submenu Linearization Points SubmenuFilter Constant Submenu Example 1 Scaling with Known Loads On-Line Calibration Input/Reading Scale and Offset MenuInput Resolution Multiplier Conversion TableExample 2 Scaling without Known Loads Page Page Enter Alarm 1 Menu Alarm 1 MenuAlarm 1 ENABLE/DISABLE Submenu Alarm 1 ABSOLUTE/DEVIATION SubmenuAlarm 1 LATCH/UNLATCH Submenu Active Submenu Contact Closure SubmenuAlarm ENABLE/DISABLE AT Power on Alarm 1 LOW Value SubmenuAlarm 1 HI Value Submenu Enter Analog Output Menu Analog Output Retransmission MenuCURRENT/VOLTAGE Submenu Analog Output ENABLE/DISABLE SubmenuReading OUTFor example Enter Alarm 2 Menu Alarm 2 MenuAlarm 2 ENABLE/DISABLE Submenu Loop Break Time MenuEnter Loop Break Time Menu Loop Break ENABLE/DISABLE Submenu Setpoint Deviation ENABLE/DISABLE SubmenuLoop Break Time Value Submenu Output 1 Menu Flow Chart for Output 1 MenuEnter Output 1 Menu Self SubmenuMINIMUM/PERCENT LOW Submenu Control Type Output MAXIMUM/PERCENT High SubmenuAction Type Submenu Auto PID SubmenuAnti Integral Submenu Proportional Band Submenu Start Auto Tune PIDReset Setup Submenu Rate Setup SubmenuCycle Time Submenu Deadband Submenu Damping Factor SubmenuOutput 2 Menu Enter Output 2 MenuControl Type Submenu Be called Reverse Acting Cycle Time Submenu Ramp ENABLE/DISABLE Submenu Ramp & Soak MenuEnter Ramp and Soak Menu Soak ENABLE/DISABLE Submenu Ramp Value SubmenuSoak Value Submenu Entering or Changing Your NON-DEFAULT ID Code ID Code MenuEnter ID Code Menu Entering or Changing Your Default ID Code Full Security Level SubmenuSETPOINT/ID Security Level Submenu Communication Option Menu 13 Flow Chart for Communication Option MenuCommunication Parameters Submenu Enter Communication Option MenuBaud Submenu Parity SubmenuBUS Format Submenu Stop BIT SubmenuModbus Protocol Submenu Line Feed SubmenuData Flow Mode Submenu Communication Interface Standard SubmenuData Separation Character Submenu Data Format Submenu Alarm Status SubmenuMain Reading Submenu Peak Value SubmenuUnit Submenu not applicable Address Setup SubmenuAddress Value Submenu Transmit Time Interval SubmenuDisplay Color Selection Menu Alarm 1 Display Color SubmenuEnter Display Color Selection Menu Normal Color Display SubmenuExample Alarm 2 Display Color SubmenuDisplay Colors change sequences Green RED AmberAmber Green Control Part SpecificationsSSR Control Output 1Excitation InsulationGeneral Factory preset value Part Factory Preset ValuesMenu Items Factory Preset Values PID Measurement Category Part CE Approvals InformationPage Page Return REQUESTS/INQUIRIES WARRANTY/DISCLAIMERPRESSURE, Strain and Force TemperatureFLOW/LEVEL Data Acquisition