Omega CNiS8DV, CNiS8DH, CNiS32, CNiS16D manual Part CE Approvals Information, Measurement Category

Page 68

PART 6

CE APPROVALS INFORMATION

This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.

Electrical Safety EN61010-1:2001

Safety requirements for electrical equipment for measurement, control and laboratory.

Double Insulation

 

Pollution Degree 2

 

Dielectric withstand Test per 1 min

 

Power to Input/Output:

2300Vac (3250Vdc)

Power to Input/Output:

1500Vac (2120Vdc)

 

(Low Voltage dc Power Option*)

 

Power to Relays/SSR Output:

2300Vac (3250Vdc)

Ethernet to Inputs:

1500Vac (2120Vdc)

Isolated RS232 to Inputs:

500Vac (720Vdc)

Isolated Analog to Inputs:

500Vac (720Vdc)

Analog/Pulse to Inputs:

No Isolation

Measurement Category I

Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 50Vac/dc. This unit should not be used in Measurement Categories II, III, IV.

Transients Overvoltage Surge (1.2 / 50uS pulse)

Input Power:

2500V

Input Power:

1500V

 

(Low Voltage dc Power Option*)

 

Ethernet:

1500V

Input/Output Signals:

500V

Note: *Units configured for external low power dc voltage, 12-36Vdc

EMC EN61326:1997 + and A1:1998 + A2:2001

Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.

EMC Emissions Table 4, Class B of EN61326

EMC Immunity** Table 1 of EN61326

Note: **I/O signal and control lines require shielded cables and these cables must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters

Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.

64

Image 68
Contents CNiS8, CNiS8C CNiS8DH, CNiS8DV CNiS16, CNiS16D, CNiS32 Servicing North America Table of Contents List of Figures List of TablesTIP Part Introduction DescriptionSafety Considerations EMC ConsiderationsBefore You Begin Part Setup Front PanelFront Panel Annunciators Rear Panel ConnectionsPower InputOutput OptionElectrical Installation Power Connections Main Power Connections Fuse RequirementsProcess Current Figure below shows the wiring hookup for 4-wire bridge input Strain Gauge11 Snubber Circuits Wiring Hookup Outputs Wiring HookupWiring Outputs RS-232 Output Wiring Hookup b RS-485 Output Wiring HookupJumper Connections Excitation OutputTop View Location of S2 on 1/8 DIN Compact Unit Part Operation Configuration Mode Turning your Controller On for the First TimeMenu Configuration Flow Chart for ID and Set Points MenuID Number Menu Entering Your NON-DEFAULT Full Security ID NumberEntering Your NON-DEFAULT SETPOINT/ID Security ID Number Set Points Menu SetpointInput Type Menu Configuration MenuEnter Configuration Menu INPUT/READING Resolution Submenu Enter Input Type MenuInput Type Menu Ratiometric Operation SubmenuReading Configuration Menu Button Selection SubmenuEnter Reading Configuration Menu Decimal Point SubmenuKNOWN/UNKNOWN Loads Submenu Linearization Points SubmenuFilter Constant Submenu Input/Reading Scale and Offset Menu Example 1 Scaling with Known Loads On-Line CalibrationExample 2 Scaling without Known Loads Input Resolution MultiplierConversion Table Page Page Alarm 1 Menu Enter Alarm 1 MenuAlarm 1 LATCH/UNLATCH Submenu Alarm 1 ENABLE/DISABLE SubmenuAlarm 1 ABSOLUTE/DEVIATION Submenu Contact Closure Submenu Active SubmenuAlarm 1 HI Value Submenu Alarm ENABLE/DISABLE AT Power onAlarm 1 LOW Value Submenu Analog Output Retransmission Menu Enter Analog Output MenuAnalog Output ENABLE/DISABLE Submenu CURRENT/VOLTAGE SubmenuReading OUTFor example Alarm 2 Menu Enter Alarm 2 MenuEnter Loop Break Time Menu Alarm 2 ENABLE/DISABLE SubmenuLoop Break Time Menu Loop Break Time Value Submenu Loop Break ENABLE/DISABLE SubmenuSetpoint Deviation ENABLE/DISABLE Submenu Flow Chart for Output 1 Menu Output 1 MenuMINIMUM/PERCENT LOW Submenu Enter Output 1 MenuSelf Submenu MAXIMUM/PERCENT High Submenu Control Type OutputAnti Integral Submenu Action Type SubmenuAuto PID Submenu Start Auto Tune PID Proportional Band SubmenuCycle Time Submenu Reset Setup SubmenuRate Setup Submenu Damping Factor Submenu Deadband SubmenuControl Type Submenu Output 2 MenuEnter Output 2 Menu Be called Reverse Acting Cycle Time Submenu Enter Ramp and Soak Menu Ramp ENABLE/DISABLE SubmenuRamp & Soak Menu Soak Value Submenu Soak ENABLE/DISABLE SubmenuRamp Value Submenu Enter ID Code Menu Entering or Changing Your NON-DEFAULT ID CodeID Code Menu SETPOINT/ID Security Level Submenu Entering or Changing Your Default ID CodeFull Security Level Submenu 13 Flow Chart for Communication Option Menu Communication Option MenuEnter Communication Option Menu Communication Parameters SubmenuBaud Submenu Parity SubmenuStop BIT Submenu BUS Format SubmenuModbus Protocol Submenu Line Feed SubmenuData Separation Character Submenu Data Flow Mode SubmenuCommunication Interface Standard Submenu Alarm Status Submenu Data Format SubmenuMain Reading Submenu Peak Value SubmenuAddress Setup Submenu Unit Submenu not applicableAddress Value Submenu Transmit Time Interval SubmenuAlarm 1 Display Color Submenu Display Color Selection MenuEnter Display Color Selection Menu Normal Color Display SubmenuAlarm 2 Display Color Submenu ExampleDisplay Colors change sequences Green RED AmberAmber Green Part Specifications ControlControl Output 1 SSRExcitation InsulationGeneral Part Factory Preset Values Factory preset valueMenu Items Factory Preset Values PID Part CE Approvals Information Measurement CategoryPage Page WARRANTY/DISCLAIMER Return REQUESTS/INQUIRIESTemperature PRESSURE, Strain and ForceFLOW/LEVEL Data Acquisition