ENTER READING CONFIGURATION MENU:
Press a
Press d
Press a
Press d
1)Press a, if necessary, until CNFG prompt appears.
2)Display advances to INPT Input Menu.
3)Display advances to RDG Reading Configuration Menu.
4)Display advances to DEC Decimal Point.
DECIMAL POINT SUBMENU:
Press d Press b
Press d
5)Display flashes previous selection for Decimal location.
6)Scroll though the available selections and choose Decimal location: FFFF, FFF.F, FF.FF or F.FFF
7)Display shows STRD stored message momentarily only, if changes were made, otherwise press a to advance to LOAD Known/Unknown Loads Submenu.
Decimal Point is passive.
KNOWN/UNKNOWN LOADS SUBMENU:
Press d
Press b
Press d
8)Display flashes previous selection of ENBL Enable or DSBL Disable.
9)Scroll though the available selection of ENBL or DSBL (flashing).
10)Display shows STRD stored message momentarily and then advances to L.PNT Linearization Points Submenu.
If ENBL Known Loads scaling method was selected, calculate the input values to the instrument based on the actual signal being received. If DSBL Without Known Loads scaling method was selected, calculate input values to the instrument based on the transducer specification.
LINEARIZATION POINTS SUBMENU:
Press d
Press b
Press d
11)Display flashes previous selection of Linearization Points Submenu.
12)Scroll though the available selections: 0002, 0003, 0004, 0006, 0006, 0007, 0008, 0009, 0010 - up to 10 Linearization Points can be selected. Default is 0002.
If display flashes NONE, your instrument has only 2 linearization points.
13)Display shows STRD stored message momentarily only, if a change was made, otherwise press a to advance to the FLTR Filter Constant Submenu.
Linearization Points allow users to customize the Transducer curve.
20