IEEE 488.2 Common Commands

 

The following table lists the IEEE 488.2 Common (*) Commands that the

 

16-Channel FET Multiplexer Modules accept. The operation of some of

 

these commands is described in Chapter 4 of this manual. For more

 

information on Common Commands, refer to the HP 75000 Series B

 

Mainframe User’s Manual (HP Model Number E1300/E1301) or the

 

ANSI/IEEE Standard 488.2-1987.

 

 

 

 

 

 

Command

Title

Description

 

 

 

 

 

 

*IDN?

Identification

Returns identification string of the switchbox.

 

 

 

*RST

Reset

Opens all channels, and invalidates current channel list for

 

 

scanning. Sets ARM:COUN 1, TRIG:SOUR IMM, and INIT:CONT OFF.

 

 

 

*TST?

Self-Test

Power-on interrupt returns +110

 

 

Incorrect MFR ID or model code returns +103

 

 

Test pass returns 0.

 

 

 

*OPC

Operation Complete

See note below.

*OPC?

Operation Complete Query

See note below.

 

 

 

*WAI

Wait to Complete

See note below.

 

 

 

*CLS

Clear status

Clears all status registers (see STATus:OPERation[:EVENt]?).

 

 

 

*ESE

Event status enable

See note below.

*ESE?

Event status enable query

See note below.

 

 

 

*ESR?

Event status register query

See note below.

 

 

 

*SRE

Service request enable

Enables status register bits (see Chapter 4 for operation).

*SRE?

Service request enable query

See note below.

 

 

 

*STB?

Read status byte query

See note below.

 

 

 

*TRG

Trigger

Triggers the switchbox to advance the scan when scan is enabled

 

 

and trigger source is TRIGger:SOURce BUS HOLD.

 

 

 

*RCL

Recall instrument state

See note below.

 

 

 

*SAV

Store instrument state

See note below.

 

 

 

Note: These commands apply to many instruments and are not documented in detail here. See the HP 75000 Series B E1300/E1301 Mainframe User’s Manual or the ANSI/IEEE Standard 488.2-1987 for more information.

76 HP E1351A/53A 16-Channel FET Multiplexer Command Reference

Chapter 5

Image 76
Contents Contents Understanding the HP E1351A/53A FET Multiplexer Modules Appendix C.HP E1351A/53A FET Multiplexer Error Messages HP E1351A/53A FET Multiplexer User’s Manual Contents Certification Documentation History July 20 According to ISO/IEC Guide 22 and ENPage Business Reply Mail Page Using This Chapter FET Multiplexer Module DescriptionAnalog Bus Connector Multimeter Connection PointsFET Multiplexer Module Block Diagram Switchbox and Scanning Voltmeter Configurations Output 70903OUTP onDigital Bus and Triggering Programming Language Initial OperationConfiguring the HP E1351A/53A FET Multiplexer Modules Card Numbers for Single-Module Switchboxes Multiplexer Card NumbersSetting the Address Switch Card Numbers for Scanning VoltmeterSetting the Interrupt Line Number Selecting the Interrupt Line NumberSetting the Card ID Switch Card ID SwitchConnecting User Inputs Channel FET Multiplexer Connector Pin-outAdding Signal Conditioning Components/Current Shunts Signal Conditioning Components/Current ShuntsConnecting Field Wiring Wiring GuidelinesWiring a Terminal Module Connecting Multimeters and Signal Generators 10. Analog Bus and Digital Bus ConnectionsAnalog Bus and Digital Bus Cables 11. Analog Bus and Digital Bus CablesPage Using the HP E1351A/53A FET Multiplexer Modules Selecting ChannelsUsing the HP E1351A/53A FET Multiplexer Modules Chapter Multiplexer Commands Matrix Commands Used in ChapterConnecting Switchbox Channels to Direct Terminals Scanport NoneCLOS? Closing Channels 02 and 09 for a Voltage Measurement Connecting Channels 02 and 10 for 4-Wire Ohms Measurement Scanning a Range of Switchbox Channels Example Making Voltage Measurements By ScanningScanning Channels 100 to 215 of a Two-Module Switchbox Example Making 2-Wire Ohms Measurements by Scanning Example Making 4-Wire Ohms Measurements by ScanningDownloading a Scan List Scanning a Switchbox without a Downloaded Scan List Output 70904INITScanning a Switchbox with a Downloaded Scan List Output 70904STATOPERENABScanning Voltmeter Configuration with HP E1326B Connecting the Analog Bus and Digital Bus CablesTemperature Measurement Using a Thermocouple Using the HP E1351A/53A FET Multiplexer Modules Chapter Using Scanning Trigger Sources Commands for Scanning Switchbox ChannelsScanning with External Instruments Scanning Commands Trigger Sources Example Scanning With External Instruments Example Scanning Using Trig Out Event In Ports Output 70914OUTP onOutput 722TRIG Extdcvmem Fifo SCANMODE, Scanport and TRIGgerSOURce Dbus Using the Scan Complete Bit Example Scan Complete MonitorCommon Command Format Command TypesScpi Command Format ROUTeSCANMODE Volt Scpi Command Reference Subsystem Syntax ABORtExample Stopping a Scan with ABORt COUNt ParametersComments Example Query Number of Scanning CyclesDISPlay Example Select Module #2 in a Switchbox for MonitoringMONitorCARD Example Enabling the Monitor Mode ModeINITiate CONTinuousExample Enabling Continuous Scans Example Query Continuous Scanning StateOUTPut STATe∙ Downloaded Scans The OUTPut subsystem does not work with Example Enabling Trig Out PortROUTe ∙ Closing Channels To close∙ Query is Software Readback The ROUTeCLOSe? command Example Closing Multiplexer ChannelsExample Query Multiplexer Channel Closure ∙ Opening Channels To open∙ Query is Software Readback The ROUTeOPEN? command ExampleOpening Multiplexer Channels Query Multiplexer Channel Open StateExample Scanning Using External Instruments ∙ Scanning Operation With a valid channel listExample Query Scan Mode Example Selecting the Abus Port Example Query Scan PortExample Setting 16 μsec SETTling Time Example Set and Query the SETTling TimeExample Enable the Status Register Bit OPERationENABleSTATus SYSTem CDEScription?∙ 16-Channel FET Multiplexer Module Description Example Reading the Description of Card #1 ModuleComments ∙ 16-Channel FET Multiplexer Module Model Number Example Reading the Model Number of a Card #1 Module∙ Executing the TRIGgerIMMediate Command a channel list Example Advancing Scan using the TRIGger CommandTRIGger IMMediate∙ Enabling the Trigger Source The TRIGgerSOURce command ∙ Trig Out Port Shared by Switchboxes See the OUTPut commandScanning Using Bus Triggers Example Scanning Using External TriggersExample Query Trigger Source Ieee 488.2 Common Commands ANSI/IEEE StandardCommand Quick Reference Scpi Commands Quick Reference DescriptionPage Maximum Voltage Screw Terminal Wire Size Watts/Slot Connectors Used P1 Maximum Offset Voltage Number of SlotsInput Impedance Power Requirements Power Off High to LowHP E1351A/53A FET Multiplexer Specifications Appendix a About This Appendix Register AddressingHP E1351A/53A Register-Based Programming Appendix B A16 Address Space Outside Command Module Or Mainframe A16 Address Space Inside Command Module Or MainframeRegister Descriptions Device Type Register Register module Register Bit DescriptionsManufacturer ID Register Enable IRQ EN Status/Control RegisterContinuous Enable CNT EN Reset Pointer RST PTR Scan Channel Delay RegisterScan Control Register Scan Channel Configuration Register Direct Channel Configuration RegisterA Tree Isolation Switch Disable A-D Valid Channel VLD C1 C0Executing Commands Direct Control RegisterROUTe ROUTeCLOSe channellist is equivalent to the following HP E1351A/53A Register-Based Programming Appendix B Register-Based Programming for Maximum Speed Decimal code Appendix B HP E1351A/53A Register-Based Programming Multimeter Command Parameter Opcodes Table B-1. Multimeter Command and Parameter OpcodesFigure B-3. Configuring the Multimeter Multimeter Register-Based Programming Error Codes Error Code CauseHP E1351A/53A FET Multiplexer Error Messages HP E1351A/53A FET Multiplexer Error Messages Appendix C Index LaddrHP E1351A/53A FET Multiplexer User’s Manual Index HP-IB HP E1351A/53A FET Multiplexer User’s Manual Index HP E1351A/53A FET Multiplexer User’s Manual Index HP E1351A/53A FET Multiplexer User’s Manual Index HP E1351A/53A FET Multiplexer User’s Manual Index