Total Organic Carbon Analyzer | Theory of Operation | |
|
|
|
Figure 2-7: Sample Handling TOC-True
2.4.2.2GAS PHASE SAMPLE HANDLING
Referencing Figure
An Oxygen Generator (P/N ST200025) is available from Teledyne that separates oxygen from the ambient air. It is a pressure swing absorption PSA device requiring only electricity only to operate. Also available from Teledyne is an Instrument Air Purifier (P/N ST200019). This device can supply the oxygen demands of the instrument using facility instrument air.
Facility carrier gas is introduced to the analyzer, regulated at 15± 2 psi, and then controlled by an internal mass flow controller
Note: Precise carrier gas flow control is mandatory for accurate
Teledyne Analytical Instruments | 13 |