Maintenance | Model 6750 |
|
|
Figure 5-11: Mass Flow Controller
5.3.10 Metering Valve (P/N ST16050)
If no O2/air flow to the sparger (P/N ST20025) is observed (no bubbles), check for tubing and fitting leaks, restrictions and adjust metering valve (be careful to return to previous indicated bubble rate). If flow cannot be restored, replace the metering valve (P/N ST16050) according to the directions below. See Figure
To replace the metering valve:
1.Turn carrier gas (Air/O2) OFF.
2.Turn pumps OFF.
3.Remove fittings and tubing.
4.Remove metering valve.
5.Recalibrate TIC if TIC and/or
6.Reinstall metering valve in reverse order.
Teledyne Analytical Instruments | 60 |