Total Organic Carbon Analyzer | Setup and Operation | |
|
|
|
Setup and Operation
Once the analyzer has been installed, it can be configured to your application. To do this you will:
•Set system parameters
•Calibrate the instrument
•Define the analysis range. Then choose
•Set alarm setpoints, and modes of alarm operation (latching, failsafe, etc.)
4.1Analyzer Startup
Referring to the Installation section and
1.Air/oxygen supply – This must provide a constant
pressure of 15 +/- 2.0 psi,
2.Drain Line – Check to see if the drain line is free of kinks and loops and allows normal vented gravity flow to an open receptacle (jar, pipe, etc.). AN AIR GAP AS DEFINED IS MANDATORY.
CAUTION: FAILURE TO OBSERVE THIS WILL RESULT IN IMPROPER OPERATION.
3.ELECTRICAL – Assure proper electrical installation. Refer to facility requirements.
4.SAFETY – All customer and Teledyne specified safety measures are followed.
Teledyne Analytical Instruments | 29 |