Total Organic Carbon Analyzer | Maintenance |
|
|
5.3 Module Service
5.3.1NDIR (Gas Calibration)
1.From the Menu, select [CALIBRATE]. The following screen will appear:
2.Select [INFRARED ANALYZER]. The following Screen will appear:
3.Disconnect tubing from reactor output and connect IR calibration adapter (P/N ST20026) as shown in Figures 5- 1,
4.Connect facility tubing to “Zero” and “Span” Gas Bottles