Performance Under Rated Conditions * - Model STR14A (0-5 to 0-500 psia)

 

Parameter

 

 

 

 

 

 

 

Description

 

Upper Range Limit **

psia

500

 

 

 

 

 

 

 

 

 

 

bar absolute

35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Minimum Span

psia

5

 

 

 

 

 

 

 

 

 

 

bar absolute

0.35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Turndown Ratio

 

 

100 to 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero Elevation and Suppression

 

No limit except minimum span from 0 to 100% URL.

 

 

 

 

 

Accuracy (Reference – Includes

 

In Analog Mode: ±0.1% of calibrated span or upper range value (URV), whichever is

 

combined effects of linearity,

 

 

greater.

 

 

 

 

 

 

hysteresis, and repeatability)

 

 

For URV below reference point (20 psi), accuracy equals:

 

 

 

 

 

Accuracy includes residual error

 

 

 

20 psi

 

1.4 bar

 

after averaging successive

 

 

±0.05 + 0.05 (

 

 

) or

±0.05 + 0.05 (

 

 

) in % of span

 

 

 

span psi

span bar

 

readings.

 

 

In Digital Mode: ±0.075% of calibrated span or upper range value (URV), whichever is

 

For FOUNDATION Fieldbus use

 

 

 

greater.

 

 

 

 

 

 

Digital Mode specifications. For

 

 

 

 

 

 

 

 

For URV below reference point (20 psi), accuracy equals:

 

HART use Analog Mode

 

 

 

specifications.

 

 

 

 

20 psi

 

 

1.4 bar

 

 

 

 

±0.025 + 0.05 (

 

) or

±0.025 + 0.05 (

 

) in % of span

 

 

 

 

span psi

span bar

*Performance specifications are based on reference conditions of 25°C (77°F), zero (0) static pressure, 10 to 55% RH, and 316L Stainless Steel barrier diaphragm.

**Transmitter URL limit or maximum seal pressure rating, whichever is lower.

Performance Under Rated Conditions – General for all Models

Parameter

Description

Output (two-wire)

Analog 4 to 20 mA or digital communications DE mode. Options available for

 

FOUNDATION Fieldbus and HART protocol.

 

 

Supply Voltage Effect

±0.005% of span per volt.

 

 

Damping Time Constant

Adjustable from 0 to 32 seconds digital damping.

 

 

RFI Protection (Standard)

Negligible (20 to 1000 MHz at 30 volts per meter).

 

 

CE Conformity (Europe)

89/336/EEC, Electromagnetic Compatibility (EMC) Directive.

 

 

NAMUR NE 43 Compliance

Transmitter failure information is generated when the measuring information is invalid or

Option

no longer present. Failure information is transmitted as a current signal but outside the

 

normal 4-20 mA measurement signal level. Transmitter failure values are: ≤ 3.6 mA and

 

≥ 21.0 mA. The normal signal range is ≥ 3.8 mA and ≤ 20.5 mA.

 

 

SIL 2/3 Compliance

SIL certified to IEC 61508 for non-redundant use in SIL 2 related Safety Systems

 

(single use) and for redundant (multiple) use in SIL 3 Safety Systems through

 

TÜV Nord Sys Tec GmbH & Co. KG under the following standards: IEC61508-1: 1998;

 

IEC 61508-2: 2000; IEC61508-3: 1998.

Image 10
Contents Guide IntroductionDescription of Diaphragm Seals 34-ST-03-64Stainless Steel Nipples are available Features DescriptionSpecifications Operating Conditions All Models Parameter Description Parameter Upper Range Limit Psig Bar Minimum Span Psig Turndown Ratio±0.10 + Or ±0.10 + Performance Under Rated Conditions General for all Models Physical and Approval Bodies 34-ST-03-64 Parameter Description 34-ST-03-64 Supply voltage/loop resistance chart Transmitter at datum Transmitter above datumTransmitter below datum Density, direct acting instrument configuration LGP/I Remote Seal With HorizontalPipe Mount DP/I Remote SealLGP/I Remote Seal With Vertical Pipe MountAP/I Remote Seal with Vertical Pipe Mount With Vertical PipeOptions Instructions Table II Seals Table II Seals Two 1/2 with plastic plugs Ansi Class 150/300/600 GFA Two 1/4 with plastic plugs NE0 Table II SealsMD0 PF0 QG0 Wetted MaterialTransmitter Mounting Brackets Options Warranty Options Table III OptionsCont’d See in Specification Dimensions and Drawings 21a Flush Flanged SealDimensions and Drawings, Dimensions and Drawings, Dimensions and Drawings, Dimensions and Drawings, This page Intentionally Blank