Honeywell STR12D warranty Density, direct acting instrument configuration

Page 16

34-ST-03-64 Page 16

Density or Interface* Calculate the minimum and maximum pressure differentials to be measured.

Pmin = (SGmin - SGf) x (d); minimum density, 4mA output

Pmax = (SGmax - SGf) x (d); maximum density, 20mA output

Where:

d = distance between the taps

SGmax = maximum Specific

Gravity

SGmin = minimum Specific

Gravity

SGf = Specific Gravity of capillary fill fluid (see page 11 for values)

*Contact STC-Phoenix concerning applications for model STR13D.

LP

HP

24257

Figure 19—Density, direct acting instrument configuration.

Image 16
Contents Guide IntroductionDescription of Diaphragm Seals 34-ST-03-64Stainless Steel Nipples are available Features DescriptionSpecifications Operating Conditions All Models Parameter Description Parameter Upper Range Limit Psig Bar Minimum Span Psig Turndown Ratio±0.10 + Or ±0.10 + Performance Under Rated Conditions General for all Models Physical and Approval Bodies 34-ST-03-64 Parameter Description 34-ST-03-64 Supply voltage/loop resistance chart Transmitter at datum Transmitter above datumTransmitter below datum Density, direct acting instrument configuration With Horizontal Pipe MountLGP/I Remote Seal DP/I Remote SealWith Vertical Pipe Mount AP/I Remote Seal with Vertical Pipe MountLGP/I Remote Seal With Vertical PipeOptions Instructions Table II Seals Table II Seals Two 1/2 with plastic plugs Ansi Class 150/300/600 GFA Two 1/4 with plastic plugs Table II Seals MD0NE0 PF0 QG0 Wetted MaterialTransmitter Mounting Brackets Options Warranty Options Table III OptionsCont’d See in Specification Dimensions and Drawings 21a Flush Flanged SealDimensions and Drawings, Dimensions and Drawings, Dimensions and Drawings, Dimensions and Drawings, This page Intentionally Blank