34-ST-03-64 Page 8
Performance Under Rated Conditions * - Model STR14G
| Parameter |
|
|
|
|
|
|
| Description | ||||
| Upper Range Limit ** | psig | 500 |
|
|
|
|
|
|
|
|
| |
|
| bar | 35 |
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Minimum Span | psig | 5 |
|
|
|
|
|
|
|
|
| |
|
| bar | 0.35 |
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Turndown Ratio |
|
| 100 to 1 |
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| ||
| Zero Elevation and Suppression |
| No limit except minimum span from absolute zero to 100% of URL. Specifications valid | ||||||||||
|
|
|
| over this range. |
|
|
|
|
| ||||
|
|
|
| ||||||||||
| Accuracy (Reference – Includes |
| In Analog Mode: ±0.1% of calibrated span or upper range value (URV), whichever is | ||||||||||
| combined effects of linearity, |
|
| greater. |
|
|
|
|
| ||||
| hysteresis, and repeatability) |
|
| For URV below reference point (20 psi), accuracy equals: | |||||||||
|
|
|
| ||||||||||
| • Accuracy includes residual error |
|
|
| 20 psi |
| 1.4 bar | ||||||
| after averaging successive |
|
| ±0.05 + 0.05 ( |
|
| ) or | ±0.05 + 0.05 ( |
|
| ) in % of span | ||
|
|
| span psi | span bar | |||||||||
| readings. |
|
| In Digital Mode: ±0.075% of calibrated span or upper range value (URV), whichever is | |||||||||
| • For FOUNDATION Fieldbus use |
| |||||||||||
|
| greater. |
|
|
|
|
| ||||||
| Digital Mode specifications. For |
|
|
|
|
|
| ||||||
|
| For URV below reference point (20 psi), accuracy equals: | |||||||||||
| HART use Analog Mode |
|
| ||||||||||
| specifications. |
|
|
|
| 20 psi |
|
| 1.4 bar | ||||
|
|
|
| ±0.025 + 0.05 ( |
| ) or | ±0.025 + 0.05 ( |
| ) in % of span | ||||
|
|
|
| span psi | span bar |
*Performance specifications are based on reference conditions of 25°C (77°F), zero (0) static pressure, 10 to 55% RH, and 316L Stainless Steel barrier diaphragm.
**Transmitter URL limit or maximum seal pressure rating, whichever is lower.