Chapter 2 Functions and Operations
Error Codes
Self Test Errors
A listing of Self Test Error messages is given in Table
Table
Error Message | Description |
BATTERY LOW Battery voltage is less than 10 volts. Charge battery. If condition per- sists, call your Anritsu Service Center.
EXTERNAL POWER LOW
PLL FAILED
External supply voltage is less than 10 volts. Call your Anritsu Service Center
INTEGRATOR | Integration circuit could not charge to a valid level. Charge battery. If |
FAILED | condition persists with a fully charged battery, call your Anritsu Ser- |
| vice Center. |
EEPROM R/W FAILED
OUT OF TEMP. RANGE
Ambient temperature is not within the specified operating range. If the temperature is within the specified operating range and the condition persists, call your Anritsu Service Center.
RTC BATTERY | The internal |
LOW | tery will affect the date stamp on saved traces. Contact your nearest |
| Anritsu Service Center. |
POWER MONITOR | Failure of the power monitor system board voltages. If condition per- |
FAIL | sists, call your Anritsu Service Center. |
BATTERY CAL | Battery communication failed. The indicated battery charge status may |
LOST | be invalid. If condition persists, call your Anritsu Service Center. |
MEMORY FAIL | The EEPROM test on the Site Master main board has failed. If condi- |
| tion persists, call your Anritsu Service Center. |