Chapter 3 Getting Started
Calibration Procedure - Single Port
Check that the “CAL OFF” message is displayed in the upper left corner of the display. This indicates that the Site Master has not been calibrated. The following procedure details how to perform the calibration.
OPEN
SHORT
LOAD
CALIBRATION
RFOUT/REFLECTION TEST PORT
TEST PORT CABLE (OPTIONAL)
Site Master S332C
|
| ESCAPE | |
1 | 2 | CLEAR | |
START | AUTO |
| |
CAL | SCALE |
| |
3 | 4 |
| |
SAVE | RECALL |
| |
SETUP | SETUP |
| |
5 | 6 |
| |
LIMIT | MARKER | ENTER | |
7 | 8 | ||
|
SAVE RECALL RUN
DISPLAY DISPLAY HOLD
9 0 +/-
ON PRINT
OFF.
MODE | FREQ/DIST | AMPLITUDE | SWEEP |
Figure 3-2. Calibration Setup
SYS
Step 1. Select the appropriate frequency range, as described in the procedure on page
Step 2. Press the START CAL key. The message CONNECT OPEN TO RF Out PORT will appear in the display.
Step 3. Connect the calibrated Open and press the ENTER key. The messages Mea- suring OPEN and then CONNECT SHORT TO RF Out PORT will appear.
Step 4. Remove the Open, connect the calibrated Short and press the ENTER key. The messages Measuring SHORT and then CONNECT TERMINATION TO RF Out will appear.
Step 5. Verify that the calibration has been properly performed by checking that the CAL ON message is now displayed in the upper left corner of the display.