ResMed 100 manual Stellar basics, About the control panel

Page 15

Stellar basics

About the control panel

l

Mains Power LED

LCD Screen

On when operating on mains power.

External power supply

LED

On when external battery is connected.

Internal battery LED

On when internal battery is in use. Flashes when device is turned off and battery is charging.

Alarm LED

Monitoring menu

Setup menu

Info menu

Red or yellow during an alarm or during alarm testing.

Alarm mute button LED

On when is pressed.

Start/stop button

Key

Start/Stop

Push dial (rotate/

click)

Therapy LED

On during treatment. Flashes during mask- fitting operation.

Function

Starts or stops treatment.

Extended hold for at least three seconds starts the mask­fitting feature.

Alarm mute

• During therapy: Press once to mute an alarm. Press a

 

 

 

second time to un-mute an alarm. If the problem is still

 

 

 

present, the alarm will sound again after two minutes.

 

 

 

See “Working with alarms” on page 13.

In stand-by mode: Extended hold for at least three seconds starts the alarm LED and alarm buzzer test.

Menu keys

Press the appropriate menu button (Monitoring, Setup, Info)

 

to enter and scroll through the relevant menu.

Push dial

Turning the dial allows you to scroll through the menu and

 

change settings. Pushing the dial allows you to enter into a

 

menu or confirm your choice.

10

Image 15
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Adverse effects IntroductionIndications for use ContraindicationsStellar at a glance ResMed USB stick Internal batteryPatient interface HumidificationSetting up for noninvasive use Remote AlarmMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Options Setup menuSetup menu Clinical Settings Mask Type Setup menu Alarm Settings Alarm VolumeSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Device Information Info menuEvent Summary Used HoursCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter Multipatient use DisinfectionServicing TroubleshootingAlarm troubleshooting Problem / possible cause Action LCD Internal Battery EmptyLCD Blocked Tube Problem / possible cause ActionLCD High Temperature 10, 11, 12 LCD Over PressureLCD Apnoea LCD Internal Battery LowLCD High Leak LCD Non-Vented MaskOther troubleshooting Device does not start when you breathe into the mask Device does not stop when you remove your maskLearn Circuit failed Technical specifications Power supply Alarm Volume RangeDimensions L x W x H Weight Air outletIEC ClassificationsAir travel requirements Page Vrms Symbols Mask should never be blocked General warnings and cautionsPage Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1 ENG