Introduction

Read the entire manual before using the device. This User Guide is for a non-clinician user, and does not contain all the information provided in the Clinical Guide.

Indications for use

The Stellar 100/150 is intended to provide ventilation for non-dependent, spontaneously breathing adult and paediatric patients (13 kg and above) with respiratory insufficiency, or respiratory failure, with or without obstructive sleep apnoea. The device is for noninvasive use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.

Contraindications

The Stellar is contraindicated in patients who are unable to endure more than brief interruptions in ventilation. The Stellar is not a life support ventilator.

If you have any of the following conditions, tell your doctor before using this device:

pneumothorax or pneumomediastinum

pathologically low blood pressure, particularly if associated with intravascular volume depletion

cerebrospinal fluid leak, recent cranial surgery or trauma

severe bullous lung disease

dehydration.

The use of the Stellar or pulse oximetry (including XPOD) is contraindicated in an MRI environment.

Adverse effects

You should report unusual chest pain, severe headache or increased breathlessness to your prescribing physician.

The following side effects may arise during the course of noninvasive ventilation with the device:

drying of the nose, mouth or throat

nosebleed

bloating

ear or sinus discomfort

eye irritation

skin rashes.

English

Introduction 1

Image 6
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Contraindications IntroductionIndications for use Adverse effectsStellar at a glance Humidification Internal batteryPatient interface ResMed USB stickSetting up for noninvasive use Remote AlarmMobile use Setting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelStarting therapy LCD screenPerforming a functional test Starting therapy Stopping therapyTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Setup menu Alarm Settings Alarm Volume Setup menuSetup menu Clinical Settings Mask Type Setup menu OptionsSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Used Hours Info menuEvent Summary Device InformationCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter Disinfection Multipatient useProblem / possible cause Action LCD Internal Battery Empty TroubleshootingAlarm troubleshooting ServicingLCD Over Pressure Problem / possible cause ActionLCD High Temperature 10, 11, 12 LCD Blocked TubeLCD Non-Vented Mask LCD Internal Battery LowLCD High Leak LCD ApnoeaOther troubleshooting Device does not start when you breathe into the mask Device does not stop when you remove your maskLearn Circuit failed Technical specifications Weight Air outlet Alarm Volume RangeDimensions L x W x H Power supplyIEC ClassificationsAir travel requirements Page Vrms Symbols General warnings and cautions Mask should never be blockedPage Limited warranty Product Warranty periodRespiratory Care Solutions Making quality of care easy EUR1 ENG