Therapy pressure

IPAP: ± 0.5 cm H2O ± 10% of set pressure (end of

tolerance

inspiration)

 

EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure

 

CPAP: ± 0.5 cm H2O ± 10% of set pressure

 

Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm

 

H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,

 

Respiratory Rate: 10 bpm, with ResMed calibration cap.

 

 

Sound pressure level

29 dBA as measured according to ISO 17510 – 1; 2002.

 

32 dBA with uncertainty of 3 dBA as measured according

 

to ISO 17510 – 1:2007.

 

 

Alarm Volume Range

> 45 dBA - <85 dBA at 1 meter (3 steps: low, medium, high)

 

 

Dimensions (L x W x H)

230 mm x 170 mm x 120 mm

 

 

Weight

2.1 kg

 

 

Air outlet

22 mm taper, compatible with ISO 5356-1:2004

 

Anaesthetic & Respiratory Equipment - Conical

 

Connectors

 

 

Pressure measurement

Internally mounted pressure transducer

 

 

Flow measurement

Internally mounted flow transducer

 

 

Power supply

AC 100–240V, 50–60Hz, 2.2 A, max. 65 W

 

 

External DC Power

24 V, 3 A

Supply (isolated)

 

 

 

Internal Battery

Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh

 

Operating hours: 2 h with a new battery under normal

 

conditions (see below).

 

Patient type: home chronic; pressure: IPAP/EPAP 15/5

 

cm H2O; mask type: Ultra Mirage; air tubing: 2 m; leak:

 

0; respiratory rate: 20 bpm; battery capacity: 100%

 

Patient type: hospital acute; pressure: IPAP/EPAP 20/5

 

cm H2O; mask type: Ultra Mirage; air tubing: 2 m; leak:

 

0; respiratory rate: 45 bpm; battery capacity: 100%

 

 

Housing construction

Flame retardant engineering thermoplastic

 

 

Environmental

• Operating temperature: 0°C to 35°C

conditions

• Operating humidity: 10%–95%non-condensing

 

• Storage and transport temperature: -20°C to 60°C

 

(+50°C*)

 

• Storage and transport humidity: 10%–95% non-

 

condensing

 

• Air pressure: 680hPa to 1,100hPa; Altitude: 3,000 m

 

*NONIN XPOD

 

 

Electromagnetic

Product complies with all applicable electromagnetic

compatibility

compatibility requirements (EMC) according to

 

IEC60601-1-2, for residential, commercial, and light

 

industry environments. For further details see “Guidance

 

and manufacturer’s declaration – electromagnetic

 

emissions and immunity” on page 30.

 

 

Air filter

Electro static fibre mesh with TPE frame structure. Bacterial

 

filtration efficiency of 99.540% on area weight 100g/m².

 

 

Air tubing

Flexible plastic, 2 m or 3 m length (22 mm diameter)

 

 

SlimLine air tubing

Flexible plastic, 1.83 m length (15 mm diameter)

 

 

28

Image 33
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Indications for use IntroductionContraindications Adverse effectsStellar at a glance Patient interface Internal batteryHumidification ResMed USB stickSetting up for noninvasive use Remote AlarmMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Clinical Settings Mask Type Setup menuSetup menu Alarm Settings Alarm Volume Setup menu OptionsSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Event Summary Info menuUsed Hours Device InformationCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter Multipatient use DisinfectionAlarm troubleshooting TroubleshootingProblem / possible cause Action LCD Internal Battery Empty ServicingLCD High Temperature 10, 11, 12 Problem / possible cause ActionLCD Over Pressure LCD Blocked TubeLCD High Leak LCD Internal Battery LowLCD Non-Vented Mask LCD ApnoeaOther troubleshooting Device does not start when you breathe into the mask Device does not stop when you remove your maskLearn Circuit failed Technical specifications Dimensions L x W x H Alarm Volume RangeWeight Air outlet Power supplyIEC ClassificationsAir travel requirements Page Vrms Symbols Mask should never be blocked General warnings and cautionsPage Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1 ENG