ResMed 100 manual Device does not start when you breathe into the mask, Learn Circuit failed

Page 31

Problem / possible cause

Solution

Treatment pressure seems high

 

 

 

Pressure required for treatment may

Consult your clinician.

have changed.

 

 

 

There is a change in the impedance in the

Perform the Learn Circuit function.

circuit configuration.

 

 

The device does not start when you breathe into the mask

 

 

SmartStart/Stop not on.

Consult your clinician.

 

 

Breath is not deep enough to trigger

Take a deep breath in and out through the

SmartStart.

mask.

 

 

There is excessive leak.

Adjust position of mask and headgear.

 

 

Plug(s) missing from port(s) on mask.

Replace plug(s).

 

 

Air tubing is not connected properly.

Connect firmly at both ends.

 

 

Air tubing is kinked or punctured.

Straighten or replace tubing.

 

 

There is a large impedance (eg,

Perform the Learn Circuit function.

antibacterial filter) in the air circuit.

 

 

The device does not stop when you remove your mask

 

 

SmartStart/Stop is disabled.

Consult your clinician.

 

 

Use of a full face mask or tracheostomy

SmartStart is disabled if Full face mask or

tube.

Trach is selected as interface.

Incompatible accessories (eg, humidifier or mask system) with high resistance being used.

Use only equipment as recommended and supplied by ResMed.

High Leak Alarm or Low Min Vent alarms

Consult your clinician.

are set to ON.

 

 

 

Confirm Stopis enabled.

Consult your clinician.

High Leak Alarm is enabled, but alarm does not activate when the mask is removed during treatment

Incompatible air delivery system being

Use only equipment as recommended

used.

and supplied by ResMed.

 

 

Pressure settings are too low for the air

Perform the Learn Circuit function to

delivery components being used.

adjust the therapy pressure according to

 

your air tubing system.

 

 

Learn Circuit failed

 

The circuit configuration is inappropriate as the impedance detected is too high.

Too many components have been included or the impedance of accessories in use is above ResMed’s recommendation, eg, type of filter, external humidifier, air tubing.

Review the components included in the circuit configuration and adjust as appropriate, then rerun Learn Circuit (see “Setup menu” on page 16).

The delivered airflow is not humid/heated although the H4i humidifier is in use

The humidifier is not properly attached.

Correctly attach the humidifier.

The humidifier does not heat.

The device is currently powered by

 

battery use or not connected to the

 

mains.

 

 

26

Image 31
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Adverse effects IntroductionIndications for use ContraindicationsStellar at a glance ResMed USB stick Internal batteryPatient interface HumidificationRemote Alarm Setting up for noninvasive useMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Attaching the pulse oximeter Working with other optional accessoriesAdding supplemental oxygen Stopping therapy using oxygen Starting therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsLCD screen Starting therapyPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Options Setup menuSetup menu Clinical Settings Mask Type Setup menu Alarm Settings Alarm VolumeParameter Description Setup menu Configuration MenuOptions On, Auto Device Information Info menuEvent Summary Used HoursData management Cleaning and maintenanceDaily Monthly WeeklyReplacing the air filter Multipatient use DisinfectionServicing TroubleshootingAlarm troubleshooting Problem / possible cause Action LCD Internal Battery EmptyLCD Blocked Tube Problem / possible cause ActionLCD High Temperature 10, 11, 12 LCD Over PressureLCD Apnoea LCD Internal Battery LowLCD High Leak LCD Non-Vented MaskOther troubleshooting Device does not stop when you remove your mask Device does not start when you breathe into the maskLearn Circuit failed Technical specifications Power supply Alarm Volume RangeDimensions L x W x H Weight Air outletClassifications IECAir travel requirements Page Vrms Symbols Mask should never be blocked General warnings and cautionsPage Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1 ENG