CAUTION
• Do not hang the air tubing in direct sunlight as the tubing may harden over time and eventually crack.
•Do not use bleach, chlorine, alcohol or
scented oils), moisturising or antibacterial soaps to clean the air tubing or the device on a daily basis (other than the approved cleaning agents Mikrozid® AF or CaviCide®). These solutions may cause hardening and reduce the life of the product. The use of cleaners and disinfecting solutions containing alcohol (other than the approved cleaning agents) is acceptable for periodic cleaning of the device, such as for cleaning between patients or especially for service periods, but not recommended for daily use.
Weekly
1Remove the air tubing from the device and the patient interface.
2Wash the air tubing in warm water using mild detergent.
3Rinse thoroughly, hang and allow to dry.
4Reconnect the air tubing to the air outlet and patient interface.
5If the device is visibly soiled, wipe the exterior surfaces of the device and the pulse oximeter (if used) with a damp cloth and mild detergent.
Monthly
1Wipe the exterior of the device and the pulse oximeter (if used) with a damp cloth and mild detergent.
2Visually inspect the air filter to check if it is blocked by dirt or contains holes.
Replacing the air filter
Replace the air filter every six months (or more often if necessary).
WARNING
Do not wash the air filter. The air filter is not washable or reusable.
1Remove the air filter cover from the back of the device.
2Remove and discard the old air filter.
3Insert a new air filter.
4Refit the air filter cover.
20