ResMed 100 manual Starting therapy, Stopping therapy

Page 17

6Check H4i heated humidifier (if in use).

Check that the warm-up feature is displayed in the Treatment screen. Start the warm-up feature. Check that the humidifier warm-up symbol is displayed on top of the screen.

Warm-up humidifier

You can use the warm-up feature to pre-heat

 

 

the water in the humidifier prior to starting

 

 

 

 

treatment. The humidifier will be automatically

 

 

detected when the device is turned on. The

 

 

Treatment screen provides the option to start

 

 

warming the humidifier. If the humidifier is

 

 

heating, the related symbol is displayed at the

 

 

top of the LCD screen.

 

 

For more information, see the H4i User Guide.

 

 

Note: The H4i in heating mode can only be used

 

 

when the device is connected to mains supply.

7Check the FiO2 monitoring sensor (if in use).

Start the FiO2 sensor calibration. Select Setup menu, then Options (see “Setup menu: Options” on page 16). Follow the instructions on the display.

8Check pulse oximeter (if in use).

Attach the accessories according to the setup descriptions (see “Attaching the pulse oximeter” on page 7). From the Monitoring menu, go to the Monitoring screen. Check that the values for SpO2 and Heart rate are displayed.

9Check oxygen connection (if in use).

Attach the accessories according to the setup descriptions (see “Adding supplemental oxygen” on page 7).

Starting therapy

CAUTION

The Clinical mode is only for clinicians. If the device is operating in clinical mode press the power switch at the back to re-start the device in patient mode .

1Fit your patient interface (mask or catheter mount) as described in the user instructions.

2Lie down and arrange the air tubing so that it is free to move if you turn in your sleep or arrange the tubing so it is comfortable in your wheelchair.

3To start treatment, press or if the SmartStart/Stop function is enabled, simply breathe into your patient interface and treatment will begin.

Stopping therapy

You can stop therapy at any time, simply remove the patient interface and press to stop airflow or if SmartStart/Stop is enabled, simply remove the patient interface, and treatment will stop automatically.

Notes:

SmartStop may not work if Full face or Trach is selected as mask type; the High Leak alarm or the Low Min Vent alarm is enabled; “Confirm Therapy Stop“ is enabled; or the mask-fitting feature is running.

When the device is stopped and operating in standby mode with an integrated humidifier connected, it will continue to blow air gently to assist cooling of the humidifier’s heater plate.

Masks with high resistance (eg, paediatric masks) may cause the operation of the SmartStop feature to be restricted.

When using with oxygen, turn off oxygen flow before stopping therapy.

12

Image 17
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Indications for use IntroductionContraindications Adverse effectsStellar at a glance Patient interface Internal batteryHumidification ResMed USB stickMobile use Setting up for noninvasive useRemote Alarm Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Adding supplemental oxygen Working with other optional accessoriesAttaching the pulse oximeter Using the FiO2 monitoring sensor Starting therapy using oxygenStopping therapy using oxygen Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsPerforming a functional test Starting therapyLCD screen Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Clinical Settings Mask Type Setup menuSetup menu Alarm Settings Alarm Volume Setup menu OptionsOptions On, Auto Setup menu Configuration MenuParameter Description Event Summary Info menuUsed Hours Device InformationDaily Cleaning and maintenanceData management Replacing the air filter WeeklyMonthly Multipatient use DisinfectionAlarm troubleshooting TroubleshootingProblem / possible cause Action LCD Internal Battery Empty ServicingLCD High Temperature 10, 11, 12 Problem / possible cause ActionLCD Over Pressure LCD Blocked TubeLCD High Leak LCD Internal Battery LowLCD Non-Vented Mask LCD ApnoeaOther troubleshooting Learn Circuit failed Device does not start when you breathe into the maskDevice does not stop when you remove your mask Technical specifications Dimensions L x W x H Alarm Volume RangeWeight Air outlet Power supplyAir travel requirements IECClassifications Page Vrms Symbols Mask should never be blocked General warnings and cautionsPage Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1 ENG