ResMed 100 manual Troubleshooting, Servicing, Alarm troubleshooting, LCD System Failure

Page 27

Servicing

CAUTION

Inspection and repair should only be performed by an authorised agent. Under no circumstances should you attempt to open, service or repair the device yourself.

This product should be inspected by an authorised ResMed service centre five years from the date of manufacture, except for the internal battery which ResMed recommends be tested after two years to assess the battery life. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed service centre.

Troubleshooting

If there is a problem, try the following suggestions. If a problem cannot be solved, contact ResMed.

Alarm troubleshooting

The most common reason for an alarm to sound is because the system has not been properly assembled. Check that the air tubing has been properly attached to the device and patient interface (and humidifier if used).

Notes:

The alarm log and alarm settings are maintained when the device is powered down and in the event of a power loss.

If multiple alarms are active simultaneously, the alarm with the highest priority will be displayed first.

If an alarm activates repeatedly, discontinue use and return the device for servicing.

Problem / possible cause

Action

 

 

LCD: Internal Battery Empty!

The remaining battery charge is below 15%. The device can be powered by the internal battery for maximum 2 minutes.

Connect the device to mains power. Note: In case of a total power failure, the therapy settings will be stored and therapy will resume when the device is powered again.

LCD: System Failure!

Component failure.

The device stops delivering air pressure (system failure 6, 7, 9, 22, 38). Therapy cannot be started (system failure 21).

1.Power-off the device.

2.Power-on the device again.

The device internal temperature is too low for the device to start/deliver therapy (system failure 21).

1.Ensure that ambient temperature is above 5°C. If the device has been stored below 5°C, allow sufficient time for it to acclimatise.

2.Power-off the device.

3.Power-on the device again.

If the problem persists, return the device for servicing.

The self test of the device fails and therapy cannot be started (system failure 21).

1.Power-off the device.

2.Power-on the device again.

If the problem persists, return the device for servicing.

22

Image 27
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Adverse effects IntroductionIndications for use ContraindicationsStellar at a glance ResMed USB stick Internal batteryPatient interface HumidificationSetting up for noninvasive use Remote AlarmMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Options Setup menuSetup menu Clinical Settings Mask Type Setup menu Alarm Settings Alarm VolumeSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Device Information Info menuEvent Summary Used HoursCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter Multipatient use DisinfectionServicing TroubleshootingAlarm troubleshooting Problem / possible cause Action LCD Internal Battery EmptyLCD Blocked Tube Problem / possible cause ActionLCD High Temperature 10, 11, 12 LCD Over PressureLCD Apnoea LCD Internal Battery LowLCD High Leak LCD Non-Vented MaskOther troubleshooting Device does not start when you breathe into the mask Device does not stop when you remove your maskLearn Circuit failed Technical specifications Power supply Alarm Volume RangeDimensions L x W x H Weight Air outletIEC ClassificationsAir travel requirements Page Vrms Symbols Mask should never be blocked General warnings and cautionsPage Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1 ENG