Parameter

Description

 

 

Learn Circuit

The device is calibrated according to your air tubing system.

 

1.

If in use, turn off the oxygen flow.

 

2.

Select the mask type.

 

3.

Set up the air circuit including accessories and patient interface.

 

 

Note: For invasive use, do not connect a catheter mount,

 

 

tracheostomy tube or HMEF, see “Setting up for invasive use”

 

 

on page 5).

 

4.

Leave the air circuit unobstructed and open to the air.

 

5.

Press

to start the Learn Circuit.

 

6.

Wait for the device to complete its automated tests (<30 sec).

 

 

The results are displayed when complete. If the circuit

 

 

configuration has been successfully learnt, displays. If

 

 

unsuccessful, displays (see “Troubleshooting” on page 22).

 

 

FiO2 Sensor

The device starts the calibration of the FiO2 monitoring sensor to

Calibration

measure the oxygen concentration of the breathable air.

 

1.

Press

to start the FiO2 sensor calibration.

 

2.

Wait for the device to complete its calibration.

The results are displayed when complete.

Note: Turn off the oxygen flow.

Ramp Time

If Max Ramp Time has been set by the clinician, you may select any

 

value up to this time.

 

Options: 0 min-Max Ramp Time (max 45 minutes, 5 minutes

 

increments)

 

 

Setup menu: Configuration Menu

Parameter

Description

 

 

Language

Sets the display language.

 

Options: Depending on regional configuration

 

 

Brightness

Sets the LCD backlight brightness.

 

Options: 20-100%, 10% increments

 

 

Backlight

Enables the LCD and keypad backlight.

 

If the AUTO setting is selected the backlight turns dark after five

 

minutes without any actions and turns on again if any button is

 

pressed again or an alarm occurs.

 

Options: On, Auto

 

 

Time format

Sets the time format.

 

Options: 24 hrs,12 hrs

 

 

Date format

Sets the date format.

 

Options: dd/mm/yyyy, mm/dd/yyyy

To return to the Options screen press the Return button

English

Setup menu

17

Image 22
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Contraindications IntroductionIndications for use Adverse effectsStellar at a glance Humidification Internal batteryPatient interface ResMed USB stickRemote Alarm Setting up for noninvasive useMobile use Setting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Attaching the pulse oximeter Working with other optional accessoriesAdding supplemental oxygen Stopping therapy using oxygen Starting therapy using oxygenUsing the FiO2 monitoring sensor Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelLCD screen Starting therapyPerforming a functional test Starting therapy Stopping therapyTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Setup menu Alarm Settings Alarm Volume Setup menuSetup menu Clinical Settings Mask Type Setup menu OptionsParameter Description Setup menu Configuration MenuOptions On, Auto Used Hours Info menuEvent Summary Device InformationData management Cleaning and maintenanceDaily Monthly WeeklyReplacing the air filter Disinfection Multipatient useProblem / possible cause Action LCD Internal Battery Empty TroubleshootingAlarm troubleshooting ServicingLCD Over Pressure Problem / possible cause ActionLCD High Temperature 10, 11, 12 LCD Blocked TubeLCD Non-Vented Mask LCD Internal Battery LowLCD High Leak LCD ApnoeaOther troubleshooting Device does not stop when you remove your mask Device does not start when you breathe into the maskLearn Circuit failed Technical specifications Weight Air outlet Alarm Volume RangeDimensions L x W x H Power supplyClassifications IECAir travel requirements Page Vrms Symbols General warnings and cautions Mask should never be blockedPage Limited warranty Product Warranty periodRespiratory Care Solutions Making quality of care easy EUR1 ENG