Problem / possible cause

Solution

The humidifier does not work.

Return the device and the humidifier for

 

servicing.

 

 

The water chamber is empty.

Fill the water chamber of the humidifier.

 

USB stick is not readable or writeable

 

 

The USB stick contains unreadable data,

Consult your clinician.

does not have enough space available, is

 

not compatible with the device.

 

 

 

USB stick is defective.

Replace the USB stick after consulting

 

your clinician.

 

 

FiO2 sensor calibration failure

 

FiO2 sensor is not attached properly.

For the correct attachment of the FiO2

 

sensor, see “Using the FiO2 monitoring

 

sensor” on page 8.

 

 

FiO2 sensor is used or defective.

If the lifetime of the FiO2 sensor has

 

exceeded one year please replace the

FiO2 sensor and start calibration again.

LCD: is displayed in the header.

 

 

Battery is not charging.

• Ensure the ambient temperature

 

 

 

conditions are within the specified

 

 

 

operating range. If the problem persists

 

 

 

within the specified operating conditions

 

 

 

please return the device for servicing.

 

 

 

• Power off the device. Power son the

 

 

 

device again.

 

 

 

Technical specifications

 

 

 

 

 

 

 

Operating pressure

• IPAP: 2 cm H2O to 40 cm H2O (in S, ST, T, PAC mode)

 

 

range

• PS: 0 cm H2O to 38 cm H2O (in S, ST, T, PAC mode)

 

 

 

• EPAP: 2 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC

 

 

 

mode)

 

 

 

• CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only)

 

 

 

• Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode)

 

 

 

• Max PS: 0 cm H2O to 30 cm H2O (in iVAPS mode)

 

 

 

Note: iVAPS mode is only available in Stellar 150.

 

 

 

 

 

 

 

Maximum single fault

60 cm H2O (in all modes)

 

 

pressure

 

 

 

 

 

 

 

 

 

Maximum breathing

2 cm H2O at 30 L/min;

 

 

resistance under single

7.2 cm H2O at 60 L/min

 

 

fault

 

 

 

 

 

 

 

 

 

 

 

Maximum flow

> 200 L/min at 20 cm H2O

 

 

Flow accuracy

± 5 L/min or 20% measured value, whichever is greater

 

 

 

 

Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm

 

 

 

H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,

 

 

 

Respiratory Rate: 10 bpm, with ResMed calibration cap.

 

 

 

 

 

English

Technical specifications

27

Image 32
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Introduction Indications for useContraindications Adverse effectsStellar at a glance Internal battery Patient interfaceHumidification ResMed USB stickMobile use Setting up for noninvasive useRemote Alarm Setting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Adding supplemental oxygen Working with other optional accessoriesAttaching the pulse oximeter Using the FiO2 monitoring sensor Starting therapy using oxygenStopping therapy using oxygen Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelPerforming a functional test Starting therapyLCD screen Starting therapy Stopping therapyTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Setup menu Setup menu Clinical Settings Mask TypeSetup menu Alarm Settings Alarm Volume Setup menu OptionsOptions On, Auto Setup menu Configuration MenuParameter Description Info menu Event SummaryUsed Hours Device InformationDaily Cleaning and maintenanceData management Replacing the air filter WeeklyMonthly Disinfection Multipatient useTroubleshooting Alarm troubleshootingProblem / possible cause Action LCD Internal Battery Empty ServicingProblem / possible cause Action LCD High Temperature 10, 11, 12LCD Over Pressure LCD Blocked TubeLCD Internal Battery Low LCD High LeakLCD Non-Vented Mask LCD ApnoeaOther troubleshooting Learn Circuit failed Device does not start when you breathe into the maskDevice does not stop when you remove your mask Technical specifications Alarm Volume Range Dimensions L x W x HWeight Air outlet Power supplyAir travel requirements IECClassifications Page Vrms Symbols General warnings and cautions Mask should never be blockedPage Limited warranty Product Warranty periodRespiratory Care Solutions Making quality of care easy EUR1 ENG