ResMed 100 manual Starting therapy, LCD screen, Performing a functional test

Page 16

LCD screen

The LCD screen displays the menus, monitoring screens and alarm status.

Values on this LCD screen and other screens throughout the User Guide are examples only.

Warm-up feature H4i

ResMed USB stick Mains power

External power supply

Internal battery

Title bar

Settings

Therapy status bar

Ramp time

Program

Therapy mode

Patient mode or Clinical mode

Current screen/number of screens within the menu

Scroll bar

Starting therapy

Performing a functional test

Make sure your device is functioning properly each time before starting therapy.

If any problems occur, see “Troubleshooting” on page 22. Please also check other provided User Instructions for troubleshooting information.

1Turn off the device by pressing the power switch at the back of the device.

2Check condition of device and accessories.

Inspect the device and all the provided accessories. If there are any visible defects, the system should not be used.

3Check the circuit configuration.

Check the integrity of the circuit configuration (device and provided accessories) according to the setup descriptions in this User Guide and that all connections are secure.

4Turn on the device and check alarms.

Press the power switch at the back of the device once to turn on the device.

Check that the alarm sounds a test beep and the LEDs (visual indicator) for the alarm signal and the Alarm mute button flash. The device is ready for use when the Treatment screen

is displayed. If the display shows the Reminder page, follow the instructions, then press

to display the Treatment screen.

5Check batteries.

Disconnect the device from the mains and external battery (if in use) so that the device is powered by the internal battery. Check that the Battery use alarm is displayed and the battery LED is on.

Note: If the charge state of the internal battery is too low, or if the battery is empty, an alarm occurs. See the Alarm troubleshooting section on page 22 for further information.

Reconnect the external battery (if in use) and check that the LED for the external power supply is lit. The External DC power use alarm will be displayed and the Alarm LED will light.

Reconnect the device to the mains.

Starting therapy

11

Image 16
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Data management Cleaning and maintenance Introduction Indications for useContraindications Adverse effectsStellar at a glance Internal battery Patient interfaceHumidification ResMed USB stickRemote Alarm Setting up for noninvasive useMobile use Setting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Attaching the pulse oximeter Working with other optional accessoriesAdding supplemental oxygen Stopping therapy using oxygen Starting therapy using oxygenUsing the FiO2 monitoring sensor Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelLCD screen Starting therapyPerforming a functional test Starting therapy Stopping therapyTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Setup menu Setup menu Clinical Settings Mask TypeSetup menu Alarm Settings Alarm Volume Setup menu OptionsParameter Description Setup menu Configuration MenuOptions On, Auto Info menu Event SummaryUsed Hours Device InformationData management Cleaning and maintenanceDaily Monthly WeeklyReplacing the air filter Disinfection Multipatient useTroubleshooting Alarm troubleshootingProblem / possible cause Action LCD Internal Battery Empty ServicingProblem / possible cause Action LCD High Temperature 10, 11, 12LCD Over Pressure LCD Blocked TubeLCD Internal Battery Low LCD High LeakLCD Non-Vented Mask LCD ApnoeaOther troubleshooting Device does not stop when you remove your mask Device does not start when you breathe into the maskLearn Circuit failed Technical specifications Alarm Volume Range Dimensions L x W x HWeight Air outlet Power supplyClassifications IECAir travel requirements Page Vrms Symbols General warnings and cautions Mask should never be blockedPage Limited warranty Product Warranty periodRespiratory Care Solutions Making quality of care easy EUR1 ENG