Results | Comparison formula | LED display | |||
LO | Displayed value < LO limit value |
|
|
| |
(Or over in the negative value) | ( Red LED on) | ||||
| |||||
OK | LO limit value ≤ Displayed value ≤ HI limit value |
|
|
| |
( Green LED on)) | |||||
|
| ||||
HI | HI limit value < Displayed value |
|
|
| |
(Or over in the positive value) | ( Red LED on) | ||||
|
LO limit value = Lower limit value
HI limit value = Upper limit value
The comparator values are common to the weighing and counting mode.
Ignore the decimal point when setting the comparator values.
For example, when the weighing capacity is 6 kg and the setting value is “001000”:
Display mode | Limit value | Capacity / Minimum display |
Normal resolution kg | 1.000 kg | 6.000 kg / 0.002 kg |
High resolution kg | 1.000 kg | 6.000 kg / 0.001 kg |
Maximum resolution kg | 0.1000 kg | 6.0000 kg / 0.0005 kg |
Normal resolution oz | 100.0 oz | 210.0 oz / 0.1 oz |
High resolution oz | 10.00 oz | 210.00 oz / 0.05 oz |
Maximum resolution oz | 10.00 oz | 210.00 oz / 0.02 oz |
Counting mode | 1000 pcs |
|
The comparator values are maintained even if the power is turned OFF.
Judgment order of comparison is from the top row to the bottom in the comparator mode tables
In this mode, the entered comparator values are not judged. Even if the upper limit value is less than the lower limit value, no error will be output.
17