12.3. Function list
Item | Parameter | Description |
| ||
Auto | | 0 | Auto | Turns the power OFF | |
|
|
|
| automatically | |
poff |
| 1 | Auto | ||
|
| ||||
Weight display |
| 0 | Normal (1/3,000) | Changes the | |
resolution |
|
|
| minimum display | |
| 1 | High (1/6,000 or 1/7,500) | |||
re5o |
| 2 | Maximum (1/12,000 or 1/15,0 00) |
| |
Weighing unit | | 0 | kg |
| |
| 1 | g |
| ||
Zero tracking |
| 0 | Zero tracking function disabled | Tracks the zero drift | |
trc | | 1 | Zero tracking function enabled |
| |
Stability band | | 0 | ±0.5d (width 1d) | Conditions to turn the | |
width |
|
|
| STABLE indicator on | |
| 1 | ±1.0d (width 2d) | |||
| 2 | ±2.0d (width 4d) | d = minimum display | ||
Stability |
| 0 | 0.5 second | ||
| in kg | ||||
detection time |
|
|
| ||
| 1 | 1.0 second |
| ||
|
| ||||
|
| 2 | 2.0 seconds |
| |
Stability |
| 0 | Weak stability / fast response | Filtering | |
Response speed |
|
|
|
| |
| 1 | Normal stability / normal response |
| ||
Cond |
| 2 | Strong stability / slow response |
| |
Serial interface | | 0 | 2400 bps | Only when the | |
baud rate |
|
|
| options, | |
| 1 | 4800 bps | |||
|
|
|
| ||
bp5 |
| 2 | 9600 bps | ||
| used. | ||||
Serial interface | | 0 | 7 bits / EVEN | ||
| |||||
data bits / parity |
|
|
|
| |
| 1 | 7 bits / ODD |
| ||
|
|
| |||
btpr |
| 2 | 8 bits / Non parity |
| |
Serial interface | | 0 | Stream mode / Command mode |
| |
Data output |
|
|
|
| |
| 1 | Command mode only |
| ||
mode |
|
|
|
| |
| 2 | Print switch mode / Command mode |
| ||
prt |
|
| |||
| 3 |
| |||
|
| 4 |
| ||
Serial interface | | 0 |
| ||
Type |
|
|
|
| |
| 1 |
| |||
5if |
| 2 |
| ||
Serial interface |
|
| ## = 00 to 99 |
| |
Address |
| ## |
| ||
|
| (Factory setting: ## = 01) |
| ||
adr |
|
|
| ||
|
|
|
| ||
Serial interface |
| 0 | No reply except data to commands |
| |
Response |
|
| |||
|
|
|
| ||
aCk | | 1 | Reply to commands |
| |
Comparator mode | | 0 |
| ||
|
| 1 |
|
23