11. CALIBRATION
Adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.
When the scale is first installed.
When the scale has been moved.
When the ambient environment has changed.
For regular calibration.
11.1. Gravity acceleration correction
When the scale is first used or has been moved to another location, it should be calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will compensate the scale. Change the gravity acceleration value stored in the scale to the value of the area where the scale will be used. Refer to the gravity acceleration map at the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be calibrated using a calibration weight at the place where it is to be used.
Make sure that the scale is in the weighing mode (displaying “kg” (“g”), “lb”, “oz” or “pcs”).
1.Press and hold the TARE switch until the gravity acceleration value (in this example: 9.7985) appears, and release the switch.
Press and hold (more than 4 seconds)
2.Press the TARE switch to enter the gravity acceleration value setting mode.
3.Change the displayed value using the following switches.
Release Press
UNITS | To shift the blinking digit to the | |
|
| right. |
|
| To increase the value of the |
| ||
|
| blinking digit by one. |
4.Press the TARE switch. The setting value is stored and “end” is displayed.
5.When calibration using a calibration weight is to be performed, go to step 3 of “11.2. Calibration using a weight”. To finish the setting procedure, turn the power OFF.
Press
Press