33
Five-level comparator modeSet the HIHI limit value.
Three-level comparator modeSet the HI limit value.
(No reply for the function setting “aCk 0 “.)
Set the six-digit value excluding the polarity and decimal point.
Command H 2 , + 0 0 0400C
RLF
Reply H 2 , + 0 0 0 4 0 0C
RLF
Five-level comparator modeSet the HI limit value.
Three-level comparator modeNot used.
(No reply for the function setting “aCk 0 “.)
Send the six-digit value excluding the polarity and decimal point.
Command H 1 , + 0 0 0300C
RLF
Reply H 1 , + 0 0 0300C
RLF
Five-level comparator modeSet the LO limit value.
Three-level comparator modeNot used.
(No reply for the function setting “aCk 0 “.)
Send the six-digit value excluding the polarity and decimal point.
Command L 1 , + 0 0 0200C
RLF
Reply L 1 , + 0 0 0200C
RLF
Five-level comparator modeSet the LOLO limit value.
Three-level comparator modeSet the LO limit value.
(No reply for the function setting “aCk 0 “.)
Send the six-digit value excluding the polarity and decimal point.
Command L 2 , + 0 0 0100C
RLF
Reply L 2 , + 0 0 0100C
RLF
Replies to the commands other than examples above when the function setting “aCk 1 “ is
selected.
The scale is not in a state where a command could be executed. Then, the scale will reply “I”.
Command Z CRLF
Reply I CRLF The scale is not in a condition that zero operation is possible.
Command does not exist for the scale. Then, the scale will reply “?”.
Command B CRLF
Reply ? CRLF The scale received an undefined command.
When the function setting “aCk 0 “ is selected, undefined commands are ignored
and no reply is sent.