15
Table 2-2. Calibration Procedure (continued)
TEST TESTED
VARIABLE
TEST POINTS TEST SEQUENCE AND ADJUSTMENTS EXPECTED
RESULTS
Power Limit
Adjust
(continued)
h. Turn A2R26 (upper knee) fully counter
clockwise. Turn on the supply and set voltage
at 200V (6010A), 20V (6011A), 60V
(6012B), 500V (6015A), and current at 5.25A
(6010A), 56A (6011A), 19A (6012B) 2.25A
(6015A), using DISPLAY SETTINGS.
i. Turn A2R26 (upper knee) clockwise until CV
LED lights. Output should be 200 ± 2V
(6010A), 20 ±0.5V (6011A), 60 ±0.4V
(6012B), and 5.25A (6010A), 55A (6011A),
18.2A (6012B), 2.2A (6015A), in CV mode.
Figure 2-1. Common Mode Setup